Product Description
Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cm) |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14either one of twoi14 |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbsi14100kgi14 |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument isI 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Advanced WVTR Testing CapabilityThis instrument is specifically designed to assess water vapor transmission rates in organosilicon coatings, ensuring high accuracy (1%) and broad measurement range (0.001-10,000 g/m224h). The digital LCD display coupled with high-precision sensors enables users to obtain reliable readings swiftly, with a response time as quick as 5 minutes. Flexible sample capacity and control modes enhance testing versatility for industrial and laboratory settings.
User-Centric Features and ComplianceThe WVTR Tester supports efficient workflows with its automatic/manual calibration, robust data logging (1000 groups), and intuitive digital interface. It's engineered to comply with global testing standards-ASTM E96, ISO 2528, and GB/T 21529-providing trusted data for international quality control. User-friendly operation, easy data output via USB or RS232, and printable reports further streamline compliance and record-keeping.
Adaptable and Robust DesignWith a sturdy stainless steel and aluminum alloy build, this tabletop instrument withstands demanding laboratory and factory environments. Test speeds (0.01-10 mm/min), controlled humidity (10-90% RH), and temperature (15-50C) support a wide array of sample and process requirements. Its automation-grade features make it suitable for repetitive, high-throughput applications while maintaining precision and reliability.
FAQ's of WVTR Testing- A Key Metric for Organosilicon Coating Quality:
Q: How does the WVTR Tester evaluate organosilicon coating quality?
A: The WVTR Tester measures the water vapor transmission rate using gravimetric and electrochemical analysis, providing quantitative data on the coating's moisture barrier performance. Accurate readings help manufacturers and researchers ensure product reliability.
Q: What types of samples and specimen sizes can be tested with this equipment?
A: This tester accommodates 1 to 3 specimens simultaneously, each with a standard size of 80 mm x 80 mm and a test width of 50 mm. The capacity is customizable to suit various research and industrial needs.
Q: When should data calibration be performed on the WVTR Tester?
A: Calibration may be conducted either automatically or manually, depending on user preference and quality protocols. Regular calibration ensures measurement accuracy and is especially recommended before critical tests or after maintenance.
Q: Where are the results and data from tests stored or exported?
A: The tester can store up to 1000 test groups in its internal memory. Results are accessible via a digital LCD display and can be exported or printed using USB or RS232 interfaces for convenient record-keeping and further analysis.
Q: What is the recommended process for conducting a WVTR test using this equipment?
A: Users install the organosilicon coating sample onto the stainless steel holder, configure the test parameters on the microcomputer control interface, and initiate the test. The device automatically regulates temperature, humidity, and pressure within compliance standards for precise measurement.
Q: How is the WVTR Tester beneficial in laboratory and industrial applications?
A: This instrument offers high accuracy, fast response, and compliance with international standards, making it integral for quality control in both research and production environments. The user-friendly interface and automation features streamline repetitive testing, enhancing productivity.
Q: Which standards does this WVTR testing equipment comply with?
A: The tester is fully compliant with ASTM E96, ISO 2528, and GB/T 21529 standards, ensuring reliable and internationally accepted test results for organosilicon coatings and related materials.