Product Details:

  • Resolution 0.001 cc/mday
  • Temperature 5C 60C
  • Humidity 5% 90% RH
  • Accuracy 1%
  • Equipment Type Oxygen Permeation Measuring System
  • Gas Pressure 0 1 MPa
  • Application Determination of oxygen transmission rate in packaging films and containers
  • Click to View more
X

Oxygen Permeation Measuring System Price And Quantity

  • 1 Set
  • 50000.00 - 100000.00 USD ($)/Set
  • 50000 USD ($)/Set

Oxygen Permeation Measuring System Product Specifications

  • 1%
  • Oxygen Permeation Measuring System
  • 5% 90% RH
  • Laboratory testing and quality control
  • 5C 60C
  • 0.001 cc/mday
  • 1/8 inch NPT
  • 100 mm x 100 mm
  • Determination of oxygen transmission rate in packaging films and containers
  • 0 1 MPa
  • 100 mm
  • High sensitivity, automatic temperature and humidity control
  • Microprocessor based digital control
  • < 2 seconds
  • 220V
  • 50 Hz
  • 650 mm
  • 3 samples
  • 1 3
  • 0 600 mm
  • 0.01 10 mm/min
  • 220V AC, 50Hz
  • 90 kg
  • Benchtop
  • 0.01 1000 cc/mday
  • 0.01 1000 cc/mday
  • Automatic
  • RS232
  • Digital LCD

Product Description

Product FeaturesNote1

Coulometric Oxygen Sensor

  • The instrument is equipped with ppb level Coulometric Oxygen Sensor of Labthink which can obtain lower test limit.
  • Designed according to ASTM D3985 with absolute value and no need to calibrate.
  • Super long service life, three times longer than traditional Coulometric oxygen sensor.
  • Functions of over limit alarm and automatic protection.

Accurate Data

  • Using 360 air circulation constant temperature technology, the temperature stability is better.
  • Equipped with high-precision temperature and humidity sensor, real-time monitoring and recording temperature and humidity changes.
  • In the test process, the automatic control of flow, temperature and relative humidity is realized, and the accuracy is higher.
  • Higher test repeatability of 0.01 cc / (m2day) can be achieved.

Effective Three-Cell

  • Independent three sets of 50cm2 standard test area cells, in line with the standard requirements for parallel sample detection.
  • Three samples are tested at the same time under the same condition, and the data are independent.
  • In the same test cycle, the number of samples completed increases from 2 to 3.
  • Automatic clamping sample, saving time and labor, clamping force is consistent, better sealing.

Intelligent Operation

  • The 12 Inch Touch panel of windows system is used for more convenient operation.
  • Automatic mode - after inputting the test temperature and humidity and clicking one key to start the test, the test proceeds automatically.
  • New drawer type test cell, one key automatic in-and-out with sound and light reminder.

Safe and Reliable

  • Safe Running: the innovative Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus, and ensure the operation reliability and data storage security.
  • Safe Operation: equipped with optical and other intelligent sensors as well as sound and light intelligent alarms to ensure the safety of operation.
  • Safe Performance: the instrument uses global famous brand with stable and reliable performance.

Powerful Functions

  • Professional test mode provides flexible and rich control functions to meet the needs of scientific research.
  • Provide oxygen permeability curve, oxygen transmission coefficient curve, temperature curve and humidity curve.
  • Ultra wide test range, meet the barrier test (customized) of various materials.
  • Ultra wide temperature range, meet the barrier test at different temperatures (customized).
  • The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen and provide oxygen free carrier gas (optional).

Testing Principle

The sample is clamped between the test chambers, oxygen or air flows on one side of the film, and high-purity nitrogen flows on the other side of the film. Oxygen molecules diffuse through the film into the high-purity nitrogen in the other side. The flowing nitrogen is carried to the sensor. Through the analysis of the oxygen concentration measured by the sensor, the oxygen transmittance is calculated.

Reference Standards

ASTM D3985, ASTM F1307, ASTM F1927(option), GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015

Test Applications

Applications
  • Films
    The oxygen permeability tests of various plastic films, paper plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film like materials.
  • Sheet
    The oxygen permeability tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.

Technical Parameters

Table 1: Test Parameters Note2

Parameters/Model C203H
Test Scope cc/(m2day) (standard area 50cm2) 0.01i200
0.005i200 (optional)
cc/(m2day) (standard area 50cm2) 0.01i400000i14Customizei14
cc/(m2day) (MASK area 5cm2) 0.12000 (optional)
cc/(m2day) (MASK area 1cm2) 0.510000 (optional)
Resolving Power cc/(m2day) 0.0001
Repeatability cc/(m2day) 0.01 or1%, whichever is greater
Temperature Range  15i50
560 (customized)
Temperature Fluctuation 0.05
Humidity Range %Rh (within standard temperature range) O2: 0%i145i901%
Carrier Gas: 0%, 5i902%
Extended Functions GP-01 Gas Purification Unit Optional
DataShieldTM Data Shield Note 3 Optional
GMP Computer System Requirements Optional
CFR21Part11 Optional

Table 2: Technical Specifications

Test Chamber 3 sets
Sample Size 4.4 x 4.4i1411.2cm11.2cmi14
Sample Thickness 120 Mili143mmi14
Standard Test Area 50cm2
Gas Specification 99.999% nitrogen, 99.5% oxygen (gas source prepared by the buyer)
Air Source Pressure 40.6 PSI / 280 kPa
Interface Size 1 / 8 "metal tube
Dimensions 23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm)
Power Supply 120VAC 10% 60Hz / 220VAC 10% 50Hz (choose one from those two)
Instrument Dimension 23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm)
Net weight 220Lbs(100kg)

Table 3: Product Configuration

Standard Configuration Mainframe, tablet computer, sampler, vacuum grease, polyurethane pipe with diameter of 6 mm
Options Gp-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements
DataShieldTM Data ShieldNote3
Remarks The compressed air inlet of the machine is I 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa); the air source is prepared by the buyer.

Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase respectively according to the usage.



Advanced Measurement Technology

This system leverages an electrochemical oxygen sensor, ensuring rapid and precise readings with a resolution of 0.001 cc/m2day and accuracy within 1%. Automatic environmental control further bolsters test consistency, while a digital LCD interface simplifies monitoring. Its microprocessor-based control streamlines test setup and execution, leading to reliable results in less than two seconds.


Efficient Data Handling and Software

Onboard memory enables secure storage for up to 10,000 test results, supporting extended laboratory workflows. The included software facilitates in-depth analysis and report generation, while USB, Ethernet, and RS232 connectivity offer streamlined data transfer and integration into lab information systems. This guarantees traceable, organized, and actionable insights for R&D and quality teams.


Easy Operation and Robust Compliance

Designed for benchtop mounting in dust-free indoor environments, the system prioritizes ease of use and operator safety. It meets strict ISO and ASTM D3985 standards and provides essential accessories for immediate operation. Onsite technical support and comprehensive documentation minimize downtime, ensuring reliable performance for manufacturers, exporters, and suppliers.

FAQ's of Oxygen Permeation Measuring System:


Q: How does the Oxygen Permeation Measuring System ensure accurate test results?

A: The system uses a standard gas calibration method in conjunction with an electrochemical oxygen sensor. Its microprocessor-based digital control automates test parameters and environmental conditions, delivering high sensitivity and accuracy (1%) within a broad measuring range.

Q: What is the process for conducting an oxygen transmission rate test with this system?

A: Operators load one to three specimens into the sample chamber, configure test settings via the digital LCD interface, and initiate the test. Automatic temperature and humidity control maintain ideal conditions. Results are swiftly generated and stored in the onboard memory for further analysis.

Q: Where is the optimal location to install and operate the measuring system?

A: The equipment is designed for benchtop mounting in indoor, dust-free laboratory environments. This setting supports consistent operation and protects the process from particulate contamination, ensuring reliable measurement outcomes.

Q: What types of samples can the Oxygen Permeation Measuring System accommodate?

A: It can test up to three samples simultaneously, accommodating specimen sizes of 100 mm x 100 mm-ideal for packaging films, foils, and containers typically used in food, pharmaceutical, or manufacturing quality control.

Q: When should calibration be performed on the system?

A: Calibration should be performed periodically using the supplied standard gas calibration standards, or when accuracy verification is required. Regular calibration assures compliance with ISO and ASTM standards and maintains data integrity.

Q: How does the dedicated analysis software add value to laboratory operations?

A: The included software streamlines test management, automates data import/export, and facilitates advanced statistical analysis. It simplifies report generation and improves traceability, reinforcing efficient and reliable laboratory workflows for quality assurance teams.

Q: What benefits does the IP20 protection class offer for the system?

A: IP20 protection ensures basic safety from finger contact and large objects, making it suitable for controlled laboratory environments. While not waterproof, it is optimized for use where environmental exposure to liquids or excessive dust is limited, further enhancing equipment longevity and reliability.

Tell us about your requirement
product

Price:

Quantity
Select Unit

  • 50
  • 100
  • 200
  • 250
  • 500
  • 1000+
Additional detail
Mobile number

Email

Other Products in 'Packaging Testing Instrument' category



Back to top
trade india member
LABTHINK INSTRUMENTS CO. LTD. All Rights Reserved.(Terms of Use)
Developed and Managed by Infocom Network Private Limited.