Product Description
Features
Professional
With innovative research and development, Labthink C640 Thickness Tester uses high precision displacement sensor, supported by scientific structure and professional control technology, which makes C640 greatly improved in stability, repeatabilityNote1 and accuracy.
- The instrument conforms to standards for mechanical contact method. The pressure foot can be lifted and lowered automatically, which can minimize the errors caused by human operation.
- Contacting time and test speed can be set as required. Customization is available for multiple contact areas and pressures, which can meet the requirements for various test conditions.
- Manual and automatic test modes can be selected by the users
- Calibration of multiple points can improve the linearity of the whole test range, which ensures the accuracy of measured results.
- Standard gauge block is available for fast calibration
High Efficiency
The instrument is featured with high efficiency and automatic design of structure, which minimize the human operation in measuring process. Intelligent controlling and data processing functions make it easy and reliable for test operation and data processing.
- Parameter program function enables the user to save all test parameters in a file so that those parameters can be applied directly for the same tests next time, which can help the user save time and minimize the error when inputting parameters manually.
- Test results include maximum value, minimum value, average value and standard variation, which can be displayed to the user directly.
- Result comparison function can help the user in evaluation of results.
- Auto sampling function (optional), feeding space, feeding speed, measure points can be set as required, which ensures the accuracy of multiple successive points measurement.
Intelligent
The instrument is embedded with Labthinks latest operating software, which is featured with user-friendly interface, intelligent data processing, strict user management and secure data storage. It also supports Labthink exclusive DataShieldTM (Optional) which provides the users with safe and reliable management of test data and test reports.
- Test data can be displayed in various forms including curves and data list
- Test data will be saved and encrypted in a unique way so that all the test information will be saved securely and reliably and protected from being tampered
- Various forms of test data can be searched, exported and printed out
Note: The described features should be in line with Table 1: Test Parameters.
Note1: The Repeatability refers to the repeatability within the test range.
Test Principle
The pre-conditioned specimen is placed on a flat test plane, the pressure foot which is paralleled to the test plane will drop down on the upper surface of the specimen with certain pressure applied on. The displacement sensor with the pressure foot will detect the space between the upper and lower surface of the specimen, which is the thickness of the specimen.
Test Standards
ISO 4593, ISO 534, ASTM D6988, ASTM F2251, GB/T 6672, GB/T 451.3, TAPPI T411, BS 2782-6, DIN 53370, ISO 3034, ISO 12625-3, ISO 5084, ASTM D374, ASTM D1777, ASTM D3652, GB/T 6547, GB/T 24218.2, FEFCO No 3, EN 1942, JIS K6250, JIS K6783, JIS Z1702
Note: The described test standards should be in line with Table 1: Test Parameters.
Applications
| Basic Applications |
- Films and Sheeting
Thickness measurement of various plastic films, sheeting, and diaphragms, etc.
|
- Paper
Thickness measurement of various paper, paper board and composite paper board, etc.
|
| Extended Applications |
- Metal Pieces & Silicon Wafers
Thickness measurement of silicon wafers, foils and various metal pieces, etc.
|
- Corrugated Paperboard
Thickness measurement of corrugated paperboard
|
- Textiles
Thickness measurement of textiles including woven fabrics, knitted fabrics, coated fabrics
|
- Non-woven Fabrics
Thickness measurement of non-woven fabrics e.g. baby diapers, sanitary napkins and medical masks, etc.
|
- Other Materials
Thickness measurement of solid insulation materials, adhesive tapes, earthwork synthetic materials and rubbers, etc.
|
Technical Specifications
Table 1: Test Parameters Note1
| Parameter Model |
C640M |
C640H |
| Test Range (Standard) |
mm |
0 2 |
0 2 |
| Resolution |
m |
0.1 |
0.1 |
| RepeatabilityNote3 |
m |
0.8 |
0.4 |
| Test Range (Optional 1) |
mm |
0 6 |
0 6 |
| Test Range (Optional 2) |
mm |
0 12 |
0 12 |
| Feeding Space |
mm |
0 1000 (Adjustable) |
| Feeding Speed |
mm/s |
1.5 80 (Adjustable) |
| Additional Functions |
Auto Sampler |
Optional |
Optional |
DataShieldTM Note2
|
Optional |
Optional |
Table 2: Technical Specifications
| Measuring Method |
Mechanical Contact Method |
| Test Pressure & Contact AreaNote3 |
Film: 17.51 KPa, 50 mm2 |
| Paper: 1001 KPa (Standard) / 501kPa (Optional), 200mm2 |
| Instrument Dimension |
370 mm (L) x 350 mm (W) x 410 mm (H) |
| Net Weight |
26kg |
Note1: The parameters in the table are measured by professional operator in Labthink laboratory according to relative requirements for laboratory standard conditions. For extended applications, customization is available to meet the special testing requirements.
Note2: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be purchase as required.
Note3: Test pressure and contact area can be selected for paper or film. There are two options for test pressure for paper, i.e. 100kPa (standard) and 50kPa(optional). For other test pressure and contact area or curved pressure foot, customizations are available.
Precision Measurement SimplifiedThe C640 Thickness Tester delivers repeatable, highly accurate measurements, thanks to its mechanical contact principle and resolution of 0.001 mm. Its semi-automatic, dual-control design ensures both flexibility and efficiency for users in laboratory and industrial environments.
Versatile Application RangeEngineered for maximum compatibility, this tester accommodates a wide selection of materials-films, papers, textiles, rubber, foils, and nonwovens. Its adjustable test range and specimen size options make it ideal for QA processes across multiple industries.
User-Friendly OperationWith a clear LCD digital display, immediate response time, and intuitive controls in both English and Chinese, the C640 Thickness Tester ensures an accessible, accurate testing experience for users of all experience levels. Dual manual and automatic modes provide operational convenience.
FAQ's of C640 Thickness Tester:
Q: How do I operate the C640 Thickness Tester for accurate measurements?
A: To ensure precise results, secure the specimen (minimum size: 4050 mm) in the test area, select manual or automatic mode, and adjust the test range as needed (0-2 mm or 0-6 mm). Use the LCD digital display to read measurements instantly, ensuring the environment is within 10C~40C and 85% relative humidity.
Q: What materials can the C640 Thickness Tester measure?
A: The C640 Thickness Tester is suitable for measuring the thickness of plastic films, paper, textiles, rubber, foils, and various nonwoven materials, making it versatile for a wide range of laboratory and industrial applications.
Q: When should I use the manual versus automatic control mode?
A: Manual mode is preferred for quick, single measurements or when operator adjustment is needed, while automatic mode is optimal for consistent, repeated testing or when enhanced throughput and uniform pressure are required.
Q: Where is the C640 Thickness Tester most effectively used?
A: This tester is most effective in laboratory settings, production quality control labs, or industrial workplaces involved in paper, film, fabric, or materials manufacturing and export, where precision thickness measurement is crucial.
Q: What is the measurement process for this thickness tester?
A: The process relies on a mechanical contact principle: the indenter with a defined contact area (50 mm2) presses against the specimen, and thickness is displayed on the LCD. The device supports fast testing (10 mm/sec) and immediate result display.
Q: How does using the C640 Thickness Tester benefit my QC operations?
A: By providing high-precision, repeatable thickness data (repeatability 0.001 mm; accuracy 0.001 mm), this tester enhances quality control, ensures compliance with material standards, and helps optimize material usage and manufacturing processes.