Product Description
Product FeaturesNote1
Multiple Test Modes & Intelligent Statistics of Qualified Specimens
- Negative pressure method
- Multiple test modes including standard, multi-grade vacuum and methylthionine chloride mode
- Automatic test with traditional methylthionine chloride
- Vacuum degree, test time and infiltration time can be adjusted and test parameters can be saved automatically so that the tests with same test parameters can be started quickly
- Automatic constant pressure compensation ensures that the tests can be performed under preset vacuum conditions
- Real-time display of test curves, easy to observe the test results
- Automatic statistics of qualified specimens
- Main components and parts are supplied by well-known global manufacturers, with guaranteed performance
Brand New Patented Intelligent Touch Operating System
- Industrial grade touch screen, one-button operation, simplified operating interface, remote upgrade and maintenance
- Chinese and English operating interface
- Measurement unit conversion
- Automatic data storage and power failure memory to prevent data loss
- Storage up to 1200 test records (standard mode)
- Multiple-level user management and login with password
- Micro-printer and USB ports for data transmission (optional)
- The instrument conforms to the requirements of GMP (optional)
- Labthinks unique DataShiledTM System for data management and connecting with information system (optional)
Test Principle
Firstly submerge the specimen in the water in the vacuum chamber, and then evacuate the vacuum chamber to form differential pressure between the inside and outside of specimen. The seal property can be obtained by observing the steady progression of bubbles from the specimen or observing how the specimen expands and restores to its original shape after vacuum release.
Test StandardNote1
ASTM D3078, GB/T 15171
Applications
| Basic Applications |
- Glass Bottles, Pipes, Cans and Boxes
|
- Plastic Bottles, Pipes, Cans and Boxes
|
- Metal Bottles, Pipes, Cans and Boxes
|
- Paper Plastic Composite Bags and Boxes
|
| Extended Applications |
|
|
|
|
|
|
Technical Specifications
| Specifications |
C660B |
| Vacuum Degree |
0 -90 KPa/ 0 -13 psi |
| Accuracy |
0.25% FS |
| Resolution |
0.1 KPa / 0.01 psi |
Vacuum Chamber Effective Sizes |
I270 mm x 210 mm (H) (standard) |
| I360 mm x 585 mm (H) (optional) |
| I460 mm x 330 mm (H) (optional) |
| Note: customization is available for other sizes |
| Gas Supply |
Air (outside of supply scope) |
| Gas Supply Pressure |
0.5 MPa 0.7 Mpa (73psi 101psi) |
| Instrument Dimension |
334 mm (L) x 230 mm (W) x 170 mm (H) |
| Power Supply |
220VAC10% 50Hz / 120VAC10% 60Hz |
| Net Weight |
Instrument: 6.5 kg Standard Vacuum Chamber: 9kg |
Configurations
- Standard Configurations
- Instrument, Standard Vacuum Chamber (I270 mm x 210 mm), I6 PU Tubing (1m)
- Optional Parts
- Micro Printer, Professional Software, Customized Vacuum Chamber, Air Compressor, GMP System, DataShieldTM Note3
- Note
- 1. The gas supply port of the instrument isI6 mm PU Tubing;
2. Customers need to prepare gas supply.
Note 1: The described test standard, applications and product features should be in line with Technical Specifications.
Note 2: The parameters in the table are measured by professional operators in Labthink laboratory under strictly controlled laboratory conditions.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Comprehensive Seal and Leak DetectionDetect leaks, vacuum failures, and compromised seals effectively in canned pet food packaging using compressed air as a test medium. This advanced instrument ensures product safety and shelf-life by providing accurate analysis in compliance with ASTM F2054 and ISO 9001 standards.
User-friendly Operation and SafetyFeaturing a digital LCD display, intuitive microcomputer control, and multilingual support (English/Chinese), this tester enhances user experience. Safety is prioritized with an emergency stop button and over-pressure alarm, making it suitable for routine use in demanding production or laboratory environments.
Data Management and FlexibilityRecord up to 1,000 test results automatically, customize test programs, and manage data efficiently via the RS232 serial interface. The device caters to various can sizes (1-5L) and ensures reliable calibration using standard weights or a force gauge for precise and repeatable outcomes.
FAQ's of Canned Pet Food Packaging Integrity Testing Instrument:
Q: How does the instrument test the integrity of canned pet food packaging?
A: The tester uses compressed air to evaluate leak, vacuum, and seal integrity. It applies pressure or vacuum and detects any packaging failures, measuring the precise force required to compromise the seal or cause leakage.
Q: What are the main benefits of using this integrity tester in a laboratory or industrial setting?
A: It guarantees accurate and repeatable test results, enhances product safety, and helps manufacturers comply with ISO 9001 and ASTM F2054 standards. The device also stores up to 1,000 test results for efficient tracking and quality control.
Q: When should calibration be performed, and how is it done?
A: Calibration should be conducted regularly or before critical testing sequences. It is performed using a standard weight or force gauge to ensure the measurement's accuracy remains within 1%.
Q: Where is data from integrity tests stored, and how can it be accessed?
A: Test results are automatically recorded in the instrument's internal memory, which can store up to 1,000 entries. Data can be accessed and managed using the intuitive digital LCD display or through the RS232 serial interface.
Q: What types and sizes of canned pet food packaging can be tested with this instrument?
A: This equipment accommodates customizable specimen sizes and is suitable for cans with capacities ranging from 1 to 5 liters, supporting testing of various standard can sizes used in the pet food industry.
Q: What safety features are included in the instrument's design?
A: Equipped with an emergency stop button and over-pressure alarm, the instrument ensures safe operation by providing immediate shut-off in critical situations. Over-load protection further enhances operator and equipment safety.
Q: How does the adjustable test speed and control mode support different testing requirements?
A: The instrument offers adjustable test speeds from 10 to 300 mm/min and utilizes microcomputer control, allowing users to tailor test parameters for different packaging types and material strengths, ensuring precise and versatile performance.