Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Reliable High-Precision TestingEnsure accurate and consistent results with high-precision humidity and temperature sensors, aided by automatic internal calibration. The stainless steel 304 chamber resists corrosion and wear, supporting long-term laboratory use in demanding environments. This ensures that every measurement aligns with ASTM, ISO, and GB/T standards.
Versatile and User-Friendly OperationOperate seamlessly with a microprocessor-based control system and intuitive LCD digital display. The included PC software enables advanced data logging and analysis, while accessories like a sample cutter and sealing ring streamline specimen preparation. Multiple sample test stations allow testing of up to three specimens at once for greater efficiency.
Comprehensive Data Management and SafetyBenefit from automatic data calculation, storage for up to 100 results, and robust safety features such as overload protection and automatic chamber lock. The tester supports RS232/USB interfaces for easy file transfer, and thermal printer output for immediate data access.
FAQ's of Water Vapor Transmission Rate Testers:
Q: How do I operate the Water Vapor Transmission Rate Tester?
A: Operation is user-friendly, featuring a microprocessor-based system with an LCD digital display. Place specimens using the included sample cutter, seal them with the provided ring, and select your test parameters. Data can be logged and analyzed via the included PC software.
Q: What materials can be tested with this equipment?
A: This WVTR tester is suitable for measuring water vapor permeability in films, sheets, foils, papers, and a variety of packaging materials, making it ideal for manufacturers and packaging suppliers.
Q: When should I use the automatic calibration feature?
A: Activate automatic calibration before starting a new set of tests or after any significant change in the testing environment, such as shifts in temperature or humidity, to ensure consistent and accurate results.
Q: Where should the instrument be placed for optimal results?
A: Place the tester in a laboratory environment with stable temperatures between 15-35C and humidity levels below 85% RH. The bench-top design facilitates easy installation in most lab settings.
Q: What is the process for logging and analyzing test data?
A: Connect the tester to a PC via RS232 or USB. The included software allows for real-time data logging, analysis, and exporting of up to 100 individual test results, with a thermal printer for quick printouts.
Q: How does the WVTR tester benefit my laboratory operations?
A: This instrument increases testing efficiency by allowing simultaneous testing of up to three specimens. It also provides precise, reliable data conforming to international standards, streamlining analysis and reporting for quality control and R&D.
Q: What safety features are integrated into the machine?
A: The tester includes overload protection and an automatic chamber lock to safeguard both the equipment and operator, ensuring safe and reliable laboratory operations.