Product Description
Product FeaturesNote1
Accurate Data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- Imported high-precision vacuum sensor can accurately test ultra-high barrier materials with 0.01i0.09cm3/ m324h0.1MPa test range.
- Imported pneumatic control system has a low failure rate and an ultra-long service life, ensuring good overall system air tightness and high test accuracy.
- Imported vacuum pump can reach 0.2Pa extreme pressure with increased vacuum rate.
- The vacuum pump is controlled automatically by the system without manual opening and closing, increases efficiency and saves labor.
- Labthinks unique high-precision automatic pressure compensation technology in the test process realizes a constant pressure difference in the high-pressure chamber, and the pressure fluctuation is less than 0.2 kPa.
- The pressure in high-pressure chamber can be set within the range of 10kPa210kPa, the pressure can be maintained accurately and stably by the system.
High Efficiency
- Six independent test cells with standard area, three times the number of test cells in traditional oxygen permeability testing instruments.
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Medium and low barrier materials, test time <4 hours (including vacuumize time).
- High barrier materials, test time <4 hours (including vacuumize time).
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click start, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with intelligent optics sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides gas transmission rate curve, gas transmission coefficient curve, and temperature curve.
- Ultra-wide test temperature range to enable the barrier tests under various extreme temperature (Optional).
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Tests of hazardous gases such as H2aCH4 are applicable (Customize).
Test Principle
The pre-conditioned specimen is mounted in the test cell as to form a sealed barrier between two chambers. The lower-pressure chamber (lower chamber) is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber (upper chamber) and a constant pressure (adjustable) difference is generated between the two chambers. The gas permeates through the specimen from higher pressure side into the lower pressure side. The gas permeability and other barrier properties of the specimen can be obtained by analyzing the pressure changes in the lower chamber.
Standard
ISO 15105-1aISO 2556aGB/T 1038-2000aASTM D1434aJIS K7126-1aYBB 00082003
Applications
| Applications |
- Films
Gas transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others
|
- Sheets
Gas transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C106H |
| Test range |
cm3/m224h0.1MPa |
0.01i50,000 |
| Resolution |
cm3/m224h0.1MPa |
0.001 |
| Test temperature |
|
15i50i145i60i14Optionali14 |
| Temperature resolution |
|
0.01 |
| Temperature fluctuation |
|
0.15 |
| Test humidity |
%RHi14Within standard test temperature rangei14 |
0%i145i902%i14Customizei14 |
| Vacuum Resolution |
Pa |
0.01 |
| Vacuum Accuracy |
Pa |
0.2%i141%-100% of Full Scalei14 |
| Vacuum Degree of Test Chamber |
Pa |
10 |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
38.48cm2 |
| Test Gas |
O2aN2aCO2 etc.i14Outside of supply scopei14 |
| Test Pressure |
10 kPa210 kPai14Can be set betweeni14 |
| Pressure Fluctuation |
0.2 kPa |
| Gas Pressure |
79.7 PSI / 550 kPa |
| Port Size |
I6 mm PU tubing |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz/220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, vacuum pump (UK), sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System requirement, DataShieldTM note3 |
| Note |
The gas compressor port of the instrument is I6 mm PU tubingi14pressure79.7 PSI/ 550kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters" "Table 1: Test Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Advanced Differential Pressure MethodThis tester employs the differential pressure method to accurately measure the gas permeability of packaging materials for heat patches. Automatic digital calibration ensures consistent precision across every test. Real-time monitoring and rapid response time streamline the testing process, making it ideal for both routine laboratory assessments and industrial quality control applications.
Comprehensive Safety and Data ManagementThe design prioritizes operator safety with overpressure protection and an emergency stop function. Data management is facilitated by storage for up to 1000 records and advanced analysis via Windows-based software. The integrated cutter aids sample preparation, while the LCD digital display and automated controls enhance usability and traceability during testing.
Versatile Material and Specimen HandlingCompatible with various materials including plastic films, composite packaging, and foils, the tester accommodates up to three specimens at a time with a maximum size of 120 120 mm. Adjustable test speed, stroke, and width allow flexibility for different sample types, supporting diverse industry needs in packaging quality assurance.
FAQ's of Gas Permeability Tester for Heat Patch Inner Packaging Quality Control:
Q: How does the Gas Permeability Tester measure the permeability of heat patch packaging?
A: The tester utilizes the differential pressure method, which measures the rate of gas transfer through packaging materials. Specimens are placed in test chambers, and the system automatically calibrates to ensure accurate and repeatable results.
Q: What types of materials can be tested with this equipment?
A: It is compatible with plastic films, composite materials, and foil packaging commonly used for heat patch inner layers, making it suitable for a wide range of laboratory and industrial packaging applications.
Q: When should I use the integrated cutter during preparation?
A: You use the integrated cutter to precisely size your specimens before loading them into the test chambers, ensuring that each sample meets the required dimensions for accurate testing.
Q: Where should the Gas Permeability Tester be installed for optimal performance?
A: For best results, install the equipment indoors in a clean, dust-free environment. Proper placement ensures stable readings and longevity of the device.
Q: What are the steps involved in the testing process?
A: Prepare the specimen using the integrated cutter, place it into one of the test chambers, set your preferred testing parameters such as speed and stroke, and start the automated process. Results are displayed digitally on the LCD and stored for analysis.
Q: How is calibration achieved and maintained?
A: Calibration is performed automatically using the machine's digital system and high-precision sensors. Multi-point calibration and temperature/humidity compensation ensure each test adheres to strict accuracy standards, reducing the need for manual adjustments.
Q: What are the advantages of using this tester for packaging quality control?
A: Benefits include high accuracy (2%), reliable safety features, fast response time (5 seconds), easy-to-use software for analysis, large data storage, and flexible specimen handling. These features enable efficient identification of packaging defects and provide comprehensive quality assurance for heat patch products.