Product Description
Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cm) |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14either one of twoi14 |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbsi14100kgi14 |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument isI 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Reliable Testing for Biodegradable PlasticsThis advanced equipment delivers comprehensive mechanical and barrier property analysis, including tensile, elongation, permeability (oxygen, carbon dioxide, and water vapor), and sealing strength. Automatic or manual calibration ensures measurement precision, while the stable benchtop design supports accurate and repeatable results. The adjustable specimen size and conditioning time expand its application for both standard and customized testing requirements.
User-Friendly Operation and Data ManagementThe analyzer features an intuitive LCD digital display and microprocessor-based control, enabling real-time data capture, storage, and output via USB or RS232. Built-in memory and PC connectivity facilitate robust data management and analysis. Multiple test modes, programmable parameters, and rapid test speeds (5-500 mm/min) maximize productivity in quality control and research settings.
Designed for Safety and EfficiencyCertified with CE marking and overload protection, the instrument prioritizes operator safety. Its design supports use in non-condensing, dust-free environments on stable, vibration-free surfaces. With an operating weight of approximately 60 kg and durable construction, it offers reliable performance in industrial and laboratory environments.
FAQ's of Mechanical and Barrier Properties Analysis of Biodegradable Plastics:
Q: How does this analyzer test the mechanical and barrier properties of biodegradable plastics?
A: The equipment uses tensile, permeability, and elongation at break principles to assess plastic films or sheets. For barrier testing, it measures how much oxygen, carbon dioxide, or water vapor passes through samples, while tensile and elongation tests determine strength and flexibility. Sealing strength assessment further ensures material integrity.
Q: What types of samples and gases can be evaluated with this equipment?
A: The analyzer is suitable for biodegradable plastic films and sheets in standard 100 x 100 mm sizes or custom dimensions. It can test permeability for gases like oxygen, CO2, water vapor, or other custom gases, depending on your application and requirements.
Q: When should I use the adjustable sample conditioning feature?
A: Adjustable sample conditioning time (12-72 hours) allows you to equilibrate samples under controlled humidity and temperature. This ensures measurement accuracy and sample consistency, particularly for comparisons or regulatory testing.
Q: Where is this analyzer optimally used?
A: Ideal for quality control labs, manufacturing facilities, and research environments, the analyzer requires a dust-free, non-condensing room with stable, vibration-free support. It is commonly used by manufacturers, exporters, and suppliers of biodegradable plastics.
Q: What is the calibration process for this instrument?
A: The analyzer provides both automatic and manual calibration options. Calibration should be performed regularly, as recommended in the manual, to maintain accuracy (0.5% FS) in all measured parameters.
Q: How is test data managed and exported from the device?
A: Test results are stored in the built-in memory and can be output in real-time or batch modes via USB or RS232. Data can be reviewed on the LCD display or transferred to a connected PC for further analysis and documentation.
Q: What are the main benefits of using this analyzer for biodegradable plastics testing?
A: Key benefits include precise multi-parameter analysis, programmable control modes, rapid data output, strong safety compliance (CE marked, overload protection), adaptable sample handling, and robust construction. These features help ensure rigorous quality standards for biodegradable plastic products.