Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Advanced Vapor Measurement TechnologyThis analyzer incorporates both gravimetric and sensor-based vapor measurement methods, providing high accuracy and consistent results. Automatic temperature and humidity control further enhance reliability, making it suitable for rigorous laboratory evaluation of PU foams and films. The equipment ensures prompt response with real-time data acquisition and a user-friendly interface.
Streamlined Sample Preparation and TestingDesigned for efficiency, the instrument features a supplied cutting die to prepare specimens to standardized dimensions. Its multi-sample capability supports simultaneous testing of up to three samples, while the variable test speed and adjustable temperature/humidity accommodate a wide range of testing conditions. Tool-free access and electronic self-check streamline maintenance.
Data Management and ComplianceThe analyzer offers secure internal memory and straightforward export of results via USB in CSV or proprietary formats. Digital microcomputer control and real-time display facilitate easy management of results. Full compliance with ISO 2528, ASTM E96, and GB/T 1037 standards ensures suitability for research and quality control environments.
FAQ's of Test machine Water Vapor Permeability Analyzer testing vapor transmission through PU foams:
Q: How does the Water Vapor Permeability Analyzer measure vapor transmission through PU foams?
A: The machine uses gravimetric and sensor-based methods to measure water vapor transmission through PU foams. Specimens are tested under controlled temperature and humidity conditions, and data is captured in real time for high accuracy.
Q: What is the process of preparing samples for testing?
A: A cutting die is provided to prepare specimens with an 85 mm diameter, ensuring consistency. Up to three samples can be loaded for simultaneous testing to increase efficiency.
Q: When should calibration be performed on the analyzer?
A: Calibration can be conducted either automatically or manually, as needed. The system allows for regular or user-initiated calibration to maintain measurement precision.
Q: Where can data from the tests be exported or accessed?
A: Test data can be stored internally and exported via USB in CSV or proprietary formats. The device also offers RS232 connectivity for flexible data management.
Q: What are the main benefits of using this analyzer for PU foam testing?
A: Key benefits include high measurement accuracy (1%), rapid response time, multi-sample testing, and compliance with international standards. The system's advanced safety, real-time monitoring, and simple maintenance add to its reliability and convenience for laboratory use.
Q: How is user safety ensured during operation?
A: The analyzer is equipped with current overload protection, gas leak detection, and automatic shutdown features in case of errors, ensuring safe operation and protecting both the user and the device.
Q: Which environmental and test parameters are adjustable during operation?
A: Operators can adjust temperature (10-65C), humidity (20-90% RH), test speed (50-500 mm/min), and other parameters via the digital touch screen for optimal testing of various PU foam and film materials.