Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Advanced Chamber Design for DurabilityConstructed from stainless steel with corrosion resistance, the analyzer's chamber guarantees long-lasting performance even under rigorous testing conditions. Its design maintains sample integrity and prevents contamination, making it ideal for repeated laboratory analyses and industrial applications alike.
Comprehensive and Flexible Testing CapabilitiesSupporting gravimetric, electrolytic, and optional infrared test modes, the analyzer accommodates a wide range of materials and requirements. Variable test speeds (0.1-10 m/min), adjustable temperature and humidity, as well as high sensitivity sensors, enable precise, tailored analysis across diverse film and packaging substrates.
User-Friendly Operation and Data ManagementEquipped with a microprocessor-based automatic control and Windows-compatible software, users benefit from digital/printed reports, intuitive touch-screen navigation, secure data storage, and convenient network connectivity via Ethernet or LAN. Quick-release sample holders and included accessories further enhance ease of use.
FAQ's of Water Vapor Permeability Analyzers:
Q: How does the Water Vapor Permeability Analyzer operate for different material types?
A: The analyzer supports multiple test modes-gravimetric, electrolytic, and optional infrared-to cover a broad spectrum of materials like films, foils, paper, and packaging. Users select the appropriate mode based on material properties, ensuring accurate and reliable water vapor transmission measurements.
Q: What is the process for calibrating the analyzer, and how often should it be done?
A: Calibration can be performed automatically or manually using the device's integrated system. Frequency of calibration depends on the intensity of use and regulatory requirements, but routine checks are advised to maintain measurement accuracy. Detailed calibration instructions are provided in the manual.
Q: When should I use the alarm functions of the analyzer?
A: The analyzer is equipped with over-temperature and sensor error alarms. These activate automatically if safe operational thresholds are exceeded or if any sensor malfunctions occur, ensuring protection of both the instrument and samples during testing.
Q: Where can I view or export my test results?
A: Test data is accessible via the digital touch screen display and can be exported as printed or digital reports. Data management is further streamlined with Windows-compatible software and can be transferred through RS232/USB or Ethernet/LAN connections to other devices or servers.
Q: What are the benefits of using a stainless steel, corrosion-resistant chamber?
A: Stainless steel with corrosion resistance offers high durability, maintains the purity of the test environment, prevents contamination, and ensures consistent performance over prolonged usage-a significant advantage in both laboratory and industrial settings.
Q: How do I prepare and secure specimens for testing?
A: Specimens, typically 80 mm or as required, are easily prepared using included sample cutters. The leak-proof, quick-release sample holders provide secure placement, reducing preparation time and minimizing any risk of leaks during tests.
Q: What makes this analyzer suitable for industrial and laboratory applications?
A: Its high sensitivity sensors, automatic control, broad operating conditions (5C-45C, 85% RH), versatile test range, and sturdy bench-top design make it ideal for environments demanding accuracy, repeatability, and ease of integration into quality control and research processes.