Product Description
Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cm) |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14either one of twoi14 |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbsi14100kgi14 |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument isI 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Superior Measurement CapabilitiesThis permeability tester offers a broad measurement range for both water vapor (0.01 - 5000 g/m224h) and oxygen (0.01 - 20000 ml/m224h0.1MPa), with a resolution down to 0.001. Its high-precision sensors and automatic calibration ensure accurate and reproducible results (2% standard deviation), supporting rigorous research and quality control in compostable film development.
Advanced Software and Data ManagementPC software enables users to customize data recording intervals, analyze test data, and easily export detailed multi-parameter reports. The selectable recording intervals (minimum 1 second) and seamless RS232/USB interface maximize compatibility and efficiency for laboratory data management.
Comprehensive User Safety and ConvenienceIntegrated safety features include over-temperature protection, automatic shutdown, and leakage detection to protect both operators and equipment. The user-friendly large LCD touch screen provides an intuitive interface for controlling test parameters, adjusting humidity and temperature, and monitoring tests in real time.
FAQ's of Testing Instrument for Water and Oxygen Transmision in Compostable Films:
Q: How does the permeability tester ensure accurate water and oxygen transmission measurements?
A: The equipment uses high-precision electrolytic and coulometric sensors, automatic zero and multi-point calibration, and programmable environmental controls, providing an accuracy of 1% of reading and reproducibility within 2% standard deviation.
Q: What types of materials and films can be tested with this instrument?
A: This tester is designed for laboratory analysis of compostable film materials, but it can also be used for other barrier films needing water vapor or oxygen transmission rate measurements, especially within the test range and specimen size options provided.
Q: When should calibration be performed on the instrument?
A: Calibration is typically required during initial setup, routine maintenance, or after any significant changes in usage conditions. The permeability tester supports automatic zero and multi-point calibration, simplifying this process for consistent results.
Q: Where should the permeability tester be installed for best performance?
A: For optimal performance and equipment longevity, install the tester indoors in a dry environment free from direct sunlight, and ensure stable room temperature and humidity conditions as recommended in the specifications.
Q: What is the process for running a transmission rate test using this device?
A: Prepare the film specimen (80 mm standard), place it in the test chamber, and use the touch screen to set the test parameters-such as temperature, humidity, and recording interval. The instrument can test 1-3 samples simultaneously, with results displayed and stored in real time.
Q: How does the integrated software contribute to test analysis and reporting?
A: The PC software allows for advanced analysis, customized report generation, export of test results, and easy adjustment of data recording intervals to match experimental needs, greatly streamlining laboratory workflows.
Q: What are the key benefits of using this permeability tester in the laboratory?
A: The main benefits include reliable and rapid test results, advanced data management and export features, user-friendly touchscreen operation, comprehensive safety protections, and flexible compatibility with various film specimens and laboratory setups.