Call Us
Send Inquiry
Measurement Instrument for Permeability of Water Vapors Through Polymer Films

Measurement Instrument for Permeability of Water Vapors Through Polymer Films

50000 USD ($)/Set

Product Details:

  • Temperature Room temperature to 50C
  • Application Determining water vapor permeation through polymer films
  • Max Height 400 mm
  • Accuracy 0.01 g/m24h
  • Resolution 0.01 g/m24h
  • Measuring Range 0.1 10,000 g/m24h
  • Automation Grade Automatic
  • Click to View more
X

Measurement Instrument for Permeability of Water Vapors Through Polymer Films Price And Quantity

  • 1 Set
  • 50000.00 - 100000.00 USD ($)/Set
  • 50000 USD ($)/Set

Measurement Instrument for Permeability of Water Vapors Through Polymer Films Product Specifications

  • 220V / 50Hz
  • Bench top
  • Standard
  • Approx. 60 kg
  • Automatic
  • USB/RS232
  • 80 mm
  • 50Hz
  • 0.01 g/m24h
  • Determining water vapor permeation through polymer films
  • 400 mm
  • Room temperature to 50C
  • Laboratory and Quality Control
  • 0.1 10,000 g/m24h
  • 0.01 g/m24h
  • Up to 100 mm
  • 1-3 specimens (simultaneous)
  • High capacity measurement
  • Digital LCD
  • Digital Control
  • 10% to 90% RH (adjustable)
  • 220V AC
  • Instant readout / 5 min
  • Structural metal housing
  • 150 mm
  • Water Vapor Permeability Tester
  • 0.1~0.6 MPa
  • Standard connection
  • 0.1 10,000 g/m24h
  • Automatic test, temperature & humidity control

Product Description

Product Featuresnote1

OTR/WVTR tested in One Instrument

  • OTR \ WVTR mode, OTR mode and WVTR mode are available for selection.
  • OTR and WVTR tests can be automatically completed for each sample mounting.
  • Genuinely reflect the OTR \ WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.

Coulometric Oxygen Sensor

  • The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
  • Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
  • Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
  • Over limit alarm and automatic protection.

Infrared Moisture Sensor

  • The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
  • Designed according to ASTM F1249.
  • Longer service life, non-consumable.
  • Over limit alarm and automatic protection.

Accurate Data

  • Using 360 air circulation constant temperature technology, the temperature stability is better.
  • Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
  • In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
  • Higher test repeatability of 0.01 cc / (m2day) can be achieved.

High Efficiency Three-cell

  • Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
  • Three samples can be tested at the same time under the same conditions, with independent tests data.
  • In the same test cycle, the number of samples completed increases from 2 to 3.
  • Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.

Intelligent Operation

  • The 12 Inch Touch panel of windows system is used for more convenient operation.
  • Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
  • New drawer type test cell, one key automatic in-and-out with sound and light reminder.

Safe and Reliable

  • Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
  • Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
  • Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.

Powerful Functions

  • Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
  • Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
  • Wide temperature range, allowing the barrier test at different temperatures (customized).
  • The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).

Testing Principle

The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.

Reference Standards

ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015

ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015

Test Applications

Applications
  • Films
    The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
  • Sheets
    The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.

Technical Specifications

Table 1: Test Parameters Note2

Parameter/Model Model C403H
OTR Test Range cc/(m2day) (Standard area 50cm2) 0.01 200
cc/(m2day) (MASK area 5cm2) 0.2 2000 (Optional)
cc/(m2day) (MASK area 1cm2) 1 10000 (Optional)
OTR Resolution cc/(m2day) 0.0001
OTR Repeatability cc/(m2day) 0.01 or 1%, whichever is greater
WVTR Test Range g/(m2day) (Standard area 50cm2) 0.005 40
g/(m2day) (MASK area 5cm2) 0.2 400 (Optional)
g/(m2day) (MASK area 1cm2) 1 2000 (Optional)
WVTR Resolution g/(m2day) 0.0001
WVTR Repeatability g/(m2day) 0.005 or 2%, whichever is greater
Temperature range C 15 - 50
5 60 (Customized)
Temperature fluctuation C 0.05
Humidity range %RH (Within standard temperature range) 0%, 5 - 902%
Extended functions GP-01 Gas Purification Unit Optional
DataShieldTM Note3 Optional
Computer System required by GMP Optional
CFR21 Part11 Optional

Table 2: Technical Specifications

Test Chamber 3 test chambers
Specimen Size 4.4 x 4.411.2cm11.2cm)
Specimen Thickness 120 Mil3mm
Standard Test Area 50 cm2
Carrier Gas 99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope)
Carrier Gas Pressure 40.6 PSI / 280 kPa
Port Size 1/8 inch metal tubing
Power Supply 120VAC10% 60Hz / 220VAC10% 50Hzeither one of two
Instrument Dimensions 23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm)
Net weight 220Lbs100kg

Configurations

Standard Configurations
Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
Optional Parts
GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
Remarks
1. The gas supply port of the instrument is 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.

Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.



Consistent and Accurate Water Vapor Analysis

Achieve highly consistent and accurate water vapor transmission rates through polymer film samples using this gravimetric method tester. With a resolution and accuracy of 0.01 g/m24h and automatic temperature and humidity control, it effortlessly delivers reliable data for research, production, and quality control needs. The factory calibration with traceable standards ensures confidence in every result.


User-Friendly Operation and Digital Control

Equipped with a digital LCD display, the instrument provides rapid, intuitive readoutsdelivering results in five minutes or less. The automatic test features and user-friendly controls simplify operation, while accessories like a sample cutting die and calibration weight streamline preparation and routine maintenance.


Rugged Construction and Comprehensive Safety

Built with structural stainless steel 304 and a sturdy bench-top design, the tester withstands rigorous laboratory use. Safety features like over-temperature and over-pressure cut-off and fuse protection guarantee secure operation. The instruments high capacity, ample reservoir, and multi-specimen capability optimize laboratory efficiency.

FAQs of Measurement Instrument for Permeability of Water Vapors Through Polymer Films:


Q: How does the Water Vapor Permeability Tester measure permeation through polymer films?

A: This instrument uses the gravimetric (desiccant) method, following ASTM E96. The polymer film specimen is mounted above a water reservoir, and the system measures the weight changes using a precision electronic balance as moisture transmits through the film.

Q: What is the process for preparing and running a test?

A: Prepare the specimen using the included cutting die, mount up to three samples in the chamber, and set the required temperature and humidity. After starting the test, the instrument automatically controls and maintains conditions, while records are displayed digitally for monitoring and export.

Q: When and how should the tester be calibrated?

A: The tester is factory calibrated with traceable standards for immediate use upon installation. Periodic calibration checks can be performed using the supplied calibration weight, ensuring sustained accuracy. Annual professional servicing is recommended for optimal performance.

Q: Where is this instrument typically used?

A: This permeability tester is suited for laboratory and quality control settings in industries like packaging, film manufacturing, and materials researchwhere accurate water vapor permeation measurement is critical.

Q: What accessories and support are provided with the tester?

A: Included are a sample cutting die, calibration weight, and instruction manual. Additionally, onsite installation and training are offered as optional services, alongside comprehensive technical support for operation and maintenance.

Q: How does the tester ensure safety during operation?

A: The device includes over-temperature and over-pressure cut-off mechanisms, along with fuse protection, to safeguard users and samples. Structural metal housing enhances safety and durability in intensive, professional environments.

Q: What are the main benefits of using this particular tester?

A: The main benefits include rapid, accurate results; the ability to test multiple specimens simultaneously; reliable digital control and data outputs; robust construction for longevity; and built-in features for service, calibration, and worker safety.

Tell us about your requirement
product

Price:

Quantity
Select Unit

  • 50
  • 100
  • 200
  • 250
  • 500
  • 1000+
Additional detail
Mobile number

Email

Other Products in 'OTR/WVTR Test System' category



Back to top
trade india member
LABTHINK INSTRUMENTS CO. LTD. All Rights Reserved.(Terms of Use)
Developed and Managed by Infocom Network Private Limited.