Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Cutting-Edge Moisture Testing for Packaging LabsPurpose-built for precision, this equipment delivers fast, sensitive moisture barrier evaluations for aluminum foil in packaging applications. The system's automatic controls, high-resolution measurements, and user-friendly interface streamline intricate lab workflows, ensuring compliance with key global standards. Its adaptability to various humidity and temperature ranges makes it suitable for diverse test conditions.
Advanced Test Chamber and Reliable Data StorageCrafted from robust stainless steel, the spacious test chamber accommodates up to three specimens, supporting a rigorous analysis of moisture permeability. The equipment's substantial data storage-10,000 results-ensures comprehensive tracking and straightforward data retrieval, critical for long-term laboratory projects and audits.
Safety, Compliance, and Seamless AutomationFeaturing overload protection, auto shut-off, and standardized calibration, this test system guarantees secure and reproducible operation. Microcomputer-controlled automation, along with USB/RS232 interfaces, facilitates integration with laboratory information management systems, enhancing productivity and data integrity for research and quality assurance teams.
FAQ's of Moisture Barrier Testing Methods for Aluminum Foil in Packaging:
Q: How does the moisture barrier testing equipment work for aluminum foil packaging?
A: The equipment evaluates the moisture permeability of aluminum foil by operating under either gravimetric or coulometric principles. It utilizes adjustable ambient air flow within a stainless steel chamber, measuring water vapor transmission through the specimen with high sensitivity and fast response, delivering digital results according to ASTM E96 and ISO 2528.
Q: What is the recommended sample preparation process before testing?
A: Samples should be prepared to the required dimensions (108 x 108 mm), ensuring surfaces are clean and free of contaminants. Proper conditioning to the specified humidity and temperature parameters is essential for accurate results during the test cycle.
Q: When should calibration be performed to ensure accurate test results?
A: Calibration should be carried out regularly using the standard reference sample provided. It is also crucial to recalibrate after significant maintenance, software updates, or changes in test conditions to maintain accuracy and compliance with standards.
Q: Where is this moisture barrier tester primarily used?
A: This equipment is mainly utilized in packaging laboratories for research, quality assurance, and materials development. Its bench-top, automated design is suited for routine and advanced analysis in industries focused on aluminum foil packaging solutions.
Q: What are the main benefits of using this testing device in packaging laboratories?
A: The device offers highly accurate, repeatable measurements, supports multi-specimen testing, ensures safety with overload protection, and boosts efficiency with automated controls and rapid digital feedback. Comprehensive data storage aids both immediate analysis and long-term record keeping.
Q: How is test data managed and accessed by users?
A: All recorded results, up to 10,000 entries, are stored internally and can be accessed or exported through a Windows-based analysis software interface via USB or RS232 port, allowing seamless integration with laboratory data management workflows.