Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Precise Testing for Medical PackagingUtilize advanced sensor technology and automated control to measure the water vapor permeability of infusion bags and other medical packaging materials with high accuracy. The tester's robust stainless steel chamber and industrial-grade construction deliver reliable results for critical quality assurance processes.
Efficient Data Management and AnalysisWith built-in internal memory for up to 500 results and included PC software, researchers and lab technicians can log, store, and analyze data seamlessly. USB and RS232 interfaces enable simple integration with laboratory networks and reporting workflows.
Safety, Compliance, and User-Friendly OperationRoHS and CE certification guarantee environmental and safety standards, while overload protection and an emergency stop increase operational safety. The digital touchscreen interface and multilingual support make it accessible for global users.
FAQ's of Infusion Bag Water Vapor Permeability Testers:
Q: How is the water vapor permeability test conducted using this equipment?
A: The infusion bag is placed in the anodized aluminum sample holder within the stainless steel chamber. The tester then measures water vapor transmission rate by controlling humidity and temperature parameters, allowing precise monitoring of water permeation through the sample.
Q: What makes the high-sensitivity humidity and temperature sensor advantageous for testing?
A: The high-sensitivity sensor ensures rapid and accurate detection of even minimal changes in humidity and temperature, improving measurement precision and enabling quick response times of one second or less, enhancing quality control reliability.
Q: When should manual or auto-calibration modes be used?
A: Manual calibration is recommended when adhering to specific laboratory protocols or when verifying accuracy against reference standards, while auto-calibration provides convenient preset adjustments for routine or frequent testing needs.
Q: Where can the Infusion Bag Water Vapor Permeability Tester be installed and operated?
A: This benchtop instrument is suitable for laboratory, research, and quality control environments. It is designed to operate safely within ambient pressures of 86 to 106 kPa and requires an AC 220V / 50Hz power source.
Q: What is the process for storing and analyzing test results?
A: Up to 500 test records can be stored in the device's internal memory. Advanced analysis can be performed using the included PC software, accessible via USB or RS232 interface, with support for English and Chinese languages.
Q: How does automated humidity and temperature control benefit the user?
A: Automated control maintains stable conditions during testing, reduces manual intervention, and ensures accuracy. Users can set desired humidity (0-99% RH) and temperature (15-60C) ranges directly on the digital touchscreen, simplifying the testing process.
Q: What safety features are integrated into the machine?
A: The tester is equipped with overload protection and an emergency stop button, minimizing risks during operation. It complies with international safety and environmental standards such as RoHS and CE for safer laboratory use.