Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Advanced Technology for Reliable Permeability AnalysisThis instrument combines state-of-the-art features such as a microcomputer control system, built-in environmental sensors, and automatic calibration to deliver precise water vapor transmission measurements. Its rapid detection capabilities and high sensitivity ensure fast and dependable results for a range of industry applications. The device is designed for accurate quality control, helping users maintain optimal packaging standards.
Easy Data Management and VersatilityEquipped with internal memory, USB export options, and a printer port, the instrument simplifies storing, exporting, and documenting test results. The included PC analysis software enables in-depth evaluation, supporting process optimization and compliance needs. Its compatibility with a wide variety of sample materials and three-sample testing capacity make it highly efficient for laboratories and production lines.
Robust Security and User-Friendly OperationSafety is prioritized with features like overload protection and an emergency stop button. The digital LCD interface and automatic data recording provide an intuitive user experience, while the bench-top design suits modern lab environments. Automatic calibration streamlines maintenance and ensures ongoing measuring accuracy.
FAQ's of Water Vapor Transmission Rate Testing Instrument for a Wide Range of Industries:
Q: How does the Water Vapor Transmission Rate Testing Instrument operate?
A: The instrument employs the gravimetric (cup) method, measuring how much water vapor permeates through sample materials under specific temperature and humidity conditions. Microcomputer control manages the test process, ensuring precise and repeatable outcomes.
Q: What materials are compatible for testing with this device?
A: It supports a broad range of materials, including plastic films, paper, coated materials, and metal foils, making it suitable for industries such as food packaging, pharmaceuticals, and electronics.
Q: When should I calibrate the instrument, and is the process automatic?
A: Calibration is automatically managed by the system, ensuring accuracy and minimizing manual intervention. Users should regularly check calibration as part of routine quality assurance procedures.
Q: Where is this instrument typically used?
A: This device is commonly used in research labs, production quality control departments, and by manufacturers and suppliers in sectors like packaging, medical, food, and electronics, especially where monitoring water vapor barrier performance is essential.
Q: What are the key benefits of using this testing instrument?
A: Benefits include fast response (within 3 seconds), high sensitivity, stable results, automatic data recording, and the capacity to test up to three specimens simultaneously, improving laboratory productivity and test reliability.
Q: How does the data management system work?
A: Test results are stored internally and can be exported via USB or RS232 ports, or printed via the printer port. Integrated PC analysis software further facilitates comprehensive result evaluation and reporting.
Q: Is the instrument straightforward to use and safe?
A: Yes, the instrument is designed for ease-of-use with an intuitive digital LCD display, emergency stop, overload protection, and a microcomputer-controlled interface ensuring both user safety and operational efficiency.