Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Precision and ComplianceDesigned in strict accordance with ASTM E96 and ISO 2528, the WVTR testing equipment delivers highly accurate and repeatable results for full-starch biodegradable film analysis in packaging. The system is equipped with a corrosion-resistant stainless steel chamber and a high-precision humidity sensor, ensuring the reliability of each test while also meeting international regulatory standards for safety and performance.
Technology and AutomationThis benchtop equipment incorporates a microprocessor-based automatic control, enabling precise adjustment of temperature (15C-55C), humidity (10-98% RH), and gas flow for superior test customization. Auto-calibration, USB/RS232 connectivity, and multilingual interface options guarantee smooth operation and adaptability to diverse laboratory workflows. Alarm functions and easy data management further enhance productivity.
Applications and BenefitsIdeal for laboratories and manufacturers, the equipment simultaneously measures up to three samples of 80 mm diameter each, streamlining the characterization of biodegradable films. Programmable test parameters and rapid response times (30 min) shorten project turnaround. The intuitive LCD display, automatic data storage, and external export capability ensure comprehensive reporting, supporting both quality assurance and innovative material research.
FAQ's of WVTR Tests for Characterization of Full-Starch Biodegradable Film for Packaging:
Q: How does the WVTR testing equipment operate for characterizing full-starch biodegradable films?
A: The equipment creates controlled environmental conditions, using automatic temperature and humidity regulation alongside dry nitrogen or air as a carrier gas. Specimens are placed in the stainless steel chamber, and the system measures water vapor transmission rates through high-precision sensors, delivering results per ASTM E96 and ISO 2528 standards.
Q: What is the typical process for conducting a WVTR test on biodegradable packaging films?
A: To perform a test, prepare film samples (80 mm diameter), mount them in the chamber, and define desired test parameters such as humidity, temperature, and gas flow. The microprocessor manages the sequence, and test results can be stored internally, exported via USB, or printed. Calibration and alerts run automatically to ensure data integrity.
Q: When should I use this WVTR testing equipment in the product development cycle?
A: The system is best used during the quality control phase and for laboratory analysis when developing or qualifying new biodegradable films. It verifies barrier properties against moisture, ensuring the materials meet packaging requirements before market release.
Q: Where is this WVTR testing machine best utilized?
A: This benchtop device is ideally suited to laboratory environments in research institutes, universities, packaging manufacturers, and quality assurance facilities, particularly where compliance with ASTM E96 and ISO 2528 is required.
Q: What benefits does simultaneous multi-specimen testing provide?
A: The ability to test up to three specimens at once improves efficiency, allowing comparative analyses and increased throughput. This is particularly valuable for laboratories tasked with processing multiple samples or carrying out batch quality checks.
Q: How does the instrument ensure data accuracy and facilitate easy reporting?
A: Accuracy is achieved through a high-precision humidity sensor (0.01 g/m224h) and auto-calibration functions. The user interface and LCD display present clear test data, while results can be easily exported via USB or printer, supporting traceable documentation.
Q: What features enhance the usability of this WVTR testing equipment?
A: Automated controls, multi-point temperature/humidity settings, alarm functions, selectable user languages (English/Chinese), and intuitive data management features all contribute to straightforward operation and minimal user intervention.