Product Description
Featuresnote1
Batch Testing
- Equipped with fully-automatic gravity fed sample feeder, which is Labthinks latest scientific and technological achievement that can facilitate continuous testing of 20 to 120 samples.
- A & B dual-channel design can test samples of two different specifications simultaneously.
- The target vacuum degree can be set by the operator to meet testing requirements of different samples.
- Dual methods of vacuum decay and pressure decay provide the flexibility to accommodate various types of package samples.
Precise Data
- Advanced pressure detection technology, using world renowned components for data stability which is not affected by ambient environment.
- Advanced microflow automatic flow control technology that can accurately simulate different sizes of leakage holes without manual adjustment.
- Both the differential pressure transducer and flow meter are traceable to NIST.
- The system can achieve a higher test repeatability of 1um.
Intelligent Control
- 15.6" embedded touch tablet computer with Windows OS.
- New longitudinal interface layout, and easy to use graphic user interface.
- Automatic alarm, automatic capture, automatic collection of unqualified samples.
- The pressure curve is displayed in real time, and the test results are counted automatically.
- Leakage rates are calculated automatically.
- The system is equipped with various sensors as intelligent reminders for safer operation and control.
- Universal printer can be connected for test results output.
- The system features embedded with USB and network ports to facilitate the external access and data transmission of the system, which can be upgraded remotely
Security Compliance
- Verified by compensation and calibration methods.
- The leak tester meets GMP requirements for data traceability and meets the needs of the pharmaceutical industry.
- User operation permission is managed at multiple levels, and the permission content can be configured on demand.
- Electronic signature is designed according to the standard requirements of 21 CFR Part11.
Test Principle
The sample is placed in the sample feeder and automatically delivered into the test cell. The leak rate and other results can be calculated and obtained by analyzing pressure changes measured by the sensor.
Reference Standards
ASTM F2338, YY-T 0681.18, and USP<1207>
Applications
| Basic Applications |
Vials |
Various vial sealing test. |
| Extended Applications |
Ampoule |
Various ampoule bottle sealing tests. |
| Cartridge Bottles |
Various cartridge bottle sealing tests. |
| Injection bottles |
Various injection bottles sealing test. |
Technical Parameters
Table 1: Test Parameters Note2
| ParameterModel |
C690H |
| Testing Range |
um(Reference aperture size USP1207) |
3i8igreat leakage |
| Detection Lower Limit |
um |
3 |
| Resolution |
um |
0.1 |
| Repeatability |
um |
1 |
| Pressure Range |
kPa |
-100i0ii14100 |
| Extended Functions |
21 CFR Part11 |
Optional |
| GMP computer system requirements |
Optional |
Table 2: Technical Specifications
| Testing Cell |
1 set for A group and 1 set for B group |
| Sample Feeder |
1 set for A group and 1 set for B group |
| Sample Size |
I45 mm80mmNote3 |
| Sample Quantity |
20120 pieces |
| Gas Specifications |
Compressed Air (Gas source is provided by the user) |
| Gas Source Pressure |
40.6 PSI / 500 kPa |
| Port Size |
I6 mm Polyurethane tube |
| Dimensions |
33.4 H x 19.6 W x 28.7 D (85cm 50cm 73cm) |
| Power |
120VAC10% 60Hz / 220VAC10% 50Hzi14select one from the twoi14 |
| Net weight |
209Lbs (95kg) |
Product Configuration
| Standard Configuration |
Mainframe, embedded tablet computer, software, flow meter, European vacuum pump, I6 mm polyurethane tube |
| Customization |
Test cell and sample feeder of group A, test cell and sample feeder of group B, negative standard reference sample and positive standard reference sample designed according to sample specifications |
| Optional Parts |
GMP computer system requirements, 21 CFR Part11, air compressor, IQ/OQ/PQ documents |
Note 1: The described product characteristics are subject to the specific annotation of the "Technical Parameters" table.
Note 2: The parameters in the table are measured in the Labthink laboratory by professional operators according to the requirements and conditions of the relevant laboratory environmental standards.
Note 3: Group C test cell can be customized for samples beyond the "Sample Size", but the lower detection limit and test range will change according to the sample size, and the actual delivery shall prevail.
Contact Us!
We can help you select or customize
Precision Leak DetectionEmploying both vacuum and pressure decay techniques, the detector provides reliable leak detection for food packaging. Its measuring range of 0-1000 mbar, with 1 mbar accuracy and 0.1 mbar resolution, ensures pinpoint identification of package leaks to maintain food safety standards.
User-Friendly OperationWith a 7-inch TFT color touchscreen and PLC control, users benefit from intuitive operation and real-time graphical analysis. The semi-automatic system facilitates consistent testing speeds from 5 to 60 cycles per minute, adapting to various workflow requirements.
Durable and Safe DesignConstructed from 304 stainless steel and fitted with an overpressure protection valve, this leak detector prioritizes safety and longevity. The chamber's silicone gasket ensures tight seals while the unit sustains over 100,000 test cycles, minimizing maintenance and disruption.
Comprehensive Data ManagementData export options include CSV and Excel files, while integrated storage securely retains up to 10,000 records. With both USB and RS232 interfaces, users can efficiently manage data and conduct post-test analysis, supporting quality assurance efforts.
FAQ's of Vacuum/Pressure Decay Food Packaging Leak Detector:
Q: How does the Vacuum/Pressure Decay Food Packaging Leak Detector identify leaks in food packaging?
A: The detector uses vacuum and pressure decay methods to identify leaks. By measuring pressure changes within the sealed chamber, it can detect even small leaks in food packages with high precision, using a measuring range of 0-1000 mbar and resolution of 0.1 mbar.
Q: What types of packaging can be tested with this device and what are the specimen size limits?
A: The device is designed for food packaging leak detection and can test specimens up to 300 300 mm in size (customizable), with pack capacities up to 5L. It accommodates various packaging types commonly used in the food industry.
Q: When should calibration be performed and how is it carried out on this equipment?
A: Calibration can be performed automatically or manually whenever accuracy checks are necessary, or as part of routine maintenance. The built-in system guides users through calibration steps to ensure reliable test results.
Q: Where can the leak detector be installed and used?
A: This machine is suitable for desktop or tabletop mounting in lab, production line, or quality control environments. Its compact design and robust construction make it ideal for manufacturers, exporters, and suppliers.
Q: What is the typical testing process and how can the speed be adjusted?
A: To conduct tests, operators place a specimen in the chamber, seal it, and activate the test using the touchscreen PLC. The test speed can be adjusted between 5 and 60 cycles per minute, allowing flexibility to match throughput needs.
Q: How does the data output feature benefit quality assurance efforts?
A: With CSV/Excel export and up to 10,000 record storage, the detector simplifies data management. Users can easily analyze test results, track historical data, and share findings, supporting comprehensive quality control and reporting.
Q: What safety features are included to protect both users and equipment?
A: Key safety features include an overpressure protection valve and high-performance silicone gasket that maintains chamber integrity. Audible and visual alarms immediately alert operators to any abnormal conditions, ensuring a secure testing environment.