Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249aISO 15106-2aGB/T 26253aJIS K7129aYBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.02i40 |
| g/(m2day) (MASK area 5cm2) |
0.2i400i14Optionali14 |
| g/(m2day) (MASK area 1cm2) |
1i2000i14Optionali14 |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard temperature rangei14 |
100%i145i902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI / 550 kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Dual Sensor AccuracyCombining Infrared and Electrolytic sensor methods, this equipment delivers precise WVTR results across a broad range (0.1 - 1000 g/m224h), with an impressive 1% accuracy. This dual-detection technology ensures reliable data for all types of films and barrier materials, supporting varied application needs in research and quality control.
Streamlined Operation & ControlThe analyzer features a user-friendly digital LCD touchscreen with microprocessor-based automatic control, promoting effortless setup and monitoring. Built-in auto-calibration and an automatic alarm system for test completion or errors enhance efficiency and reduce the risk of operator error. Data is easily exportable in CSV and PDF formats via USB, RS232, or Ethernet.
High Throughput & Flexible TestingThis system allows simultaneous testing of up to three specimens, managing sample sizes of 80 mm x 80 mm, and up to 5 mm thick. Its rapid response time (under 10 minutes), wide test stroke (10-50 mm), and high storage capacity (10,000 results) support intensive laboratory and industrial QC workflows, maximizing productivity per day.
FAQ's of Rapid WVTR Testing with Infrared & Electrolytic Sensor Methods:
Q: How does the Rapid WVTR Testing System ensure accurate moisture vapor transmission rate measurements?
A: It utilizes a dual sensor system-Infrared and Electrolytic-which enables precise detection of water vapor across a wide measurement range. Coupled with built-in auto-calibration, this ensures results are both rapid and highly accurate, with an accuracy rate of 1% of reading.
Q: What is the typical process for conducting a WVTR test with this analyzer?
A: Simply prepare your samples (up to three at once), configure the relevant test parameters on the touchscreen interface, and start the test. The analyzer automatically manages control, calibration, and data acquisition. Results are displayed digitally and can be saved or exported instantly.
Q: When is the best environment to operate this equipment?
A: Optimal operation requires a dust-free and vibration-free setup. Environmental requirements include adjustable humidity (0-100% RH) and temperature (15C to 60C) to accommodate a variety of testing standards and ensure measurement reliability.
Q: Where can test data from the analyzer be accessed and exported?
A: Test results are digitally displayed and can be stored within the instrument's internal memory (up to 10,000 results). Data can be exported in CSV or PDF format via USB, RS232, or Ethernet interfaces, facilitating easy integration with laboratory information systems or reports.
Q: What materials and sample types are compatible with the Rapid WVTR Testing System?
A: The device is designed for films, foils, and barrier materials used in packaging and industrial applications. Standard specimen dimensions are 80 mm x 80 mm, with thicknesses up to 5 mm, making it versatile for diverse lab and industrial requirements.
Q: How can users benefit from the automatic alarm and calibration features?
A: Automatic alarms alert operators when tests are complete or if errors occur, minimizing oversight and ensuring process integrity. The built-in auto-calibration feature guarantees reliable results and reduces maintenance or manual calibration workloads.
Q: What distinguishes this equipment in terms of speed and capacity?
A: With a rapid response time (under 10 minutes) and high throughput (up to 20 tests per day), this analyzer enables efficient and continuous workflow in quality control or research environments. The simultaneous multi-specimen capability further boosts productivity.