Product Description
Product FeaturesNote1
Coulometric Oxygen Sensor
- The instrument is equipped with ppb level Coulometric Oxygen Sensor of Labthink which can obtain lower test limit.
- Designed according to ASTM D3985 with absolute value and no need to calibrate.
- Super long service life, three times longer than traditional Coulometric oxygen sensor.
- Functions of over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, real-time monitoring and recording temperature and humidity changes.
- In the test process, the automatic control of flow, temperature and relative humidity is realized, and the accuracy is higher.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
Effective Three-Cell
- Independent three sets of 50cm2 standard test area cells, in line with the standard requirements for parallel sample detection.
- Three samples are tested at the same time under the same condition, and the data are independent.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic clamping sample, saving time and labor, clamping force is consistent, better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting the test temperature and humidity and clicking one key to start the test, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: the innovative Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus, and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as sound and light intelligent alarms to ensure the safety of operation.
- Safe Performance: the instrument uses global famous brand with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and rich control functions to meet the needs of scientific research.
- Provide oxygen permeability curve, oxygen transmission coefficient curve, temperature curve and humidity curve.
- Ultra wide test range, meet the barrier test (customized) of various materials.
- Ultra wide temperature range, meet the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen and provide oxygen free carrier gas (optional).
Testing Principle
The sample is clamped between the test chambers, oxygen or air flows on one side of the film, and high-purity nitrogen flows on the other side of the film. Oxygen molecules diffuse through the film into the high-purity nitrogen in the other side. The flowing nitrogen is carried to the sensor. Through the analysis of the oxygen concentration measured by the sensor, the oxygen transmittance is calculated.
Reference Standards
ASTM D3985, ASTM F1307, ASTM F1927(option), GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
Test Applications
| Applications |
- Films
The oxygen permeability tests of various plastic films, paper plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film like materials.
|
- Sheet
The oxygen permeability tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Parameters
Table 1: Test Parameters Note2
| Parameters/Model |
C203H |
| Test Scope |
cc/(m2day) (standard area 50cm2) |
0.01i200 0.005i200 (optional) |
| cc/(m2day) (standard area 50cm2) |
0.01i400000i14Customizei14 |
| cc/(m2day) (MASK area 5cm2) |
0.12000 (optional) |
| cc/(m2day) (MASK area 1cm2) |
0.510000 (optional) |
| Resolving Power |
cc/(m2day) |
0.0001 |
| Repeatability |
cc/(m2day) |
0.01 or1%, whichever is greater |
| Temperature Range |
|
15i50 560 (customized) |
| Temperature Fluctuation |
|
0.05 |
| Humidity Range |
%Rh (within standard temperature range) |
O2: 0%i145i901% Carrier Gas: 0%, 5i902% |
| Extended Functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Data Shield Note 3 |
Optional |
| GMP Computer System Requirements |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 sets |
| Sample Size |
4.4 x 4.4i1411.2cm11.2cmi14 |
| Sample Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50cm2 |
| Gas Specification |
99.999% nitrogen, 99.5% oxygen (gas source prepared by the buyer) |
| Air Source Pressure |
40.6 PSI / 280 kPa |
| Interface Size |
1 / 8 "metal tube |
| Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Power Supply |
120VAC 10% 60Hz / 220VAC 10% 50Hz (choose one from those two) |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Net weight |
220Lbs(100kg) |
Table 3: Product Configuration
| Standard Configuration |
Mainframe, tablet computer, sampler, vacuum grease, polyurethane pipe with diameter of 6 mm |
| Options |
Gp-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements DataShieldTM Data ShieldNote3 |
| Remarks |
The compressed air inlet of the machine is I 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa); the air source is prepared by the buyer. |
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase respectively according to the usage.
Ensuring Oxygen Barrier Performance for Edible Oil PackagingWith a measurement range of 0.01-1000 cc/m2day, this OTR test instrument provides precise and reliable assessment of oxygen impermeability in packaging films and containers. The pre-conditioning chamber and temperature/humidity control simulate real-world conditions, ensuring results reflect actual performance. Whether you're manufacturing or exporting edible oil packaging, this tool helps prevent oxidation and spoilage.
User-Centric and Automated Testing WorkflowFeaturing an intuitive multilingual LCD touch interface and fully automatic operations, the system simplifies the OTR measurement process. Data logging, multi-user access, and auto-calibration ensure both ease of use and data integrity. Audible and visual alarms alert users when limits are exceeded, while comprehensive software enables advanced analyses and secure data management.
FAQ's of Essential Tests for Selecting Effective Edible Oil Packaging - Oxygen Transmission Rate (OTR) Test:
Q: How does the OTR test instrument assess the suitability of edible oil packaging?
A: The instrument measures the oxygen transmission rate using the coulometric sensor method under controlled temperature and humidity, as per ASTM D3985 and ISO 15105-2 standards. This reveals how much oxygen passes through the packaging, helping determine if it provides adequate protection for edible oils.
Q: What is the typical process for conducting an OTR test on edible oil packaging?
A: Specimens are first pre-conditioned in the provided chamber, then placed in the instrument. The system automatically controls conditions, initiates testing, and records oxygen transmission data, displaying results digitally. Data can be exported in CSV or PDF format for analysis or compliance reporting.
Q: When should manufacturers or exporters utilize this OTR test?
A: OTR testing is essential during product development, quality control, and prior to exporting edible oil packaging to ensure compliance with international standards, as well as to optimize barrier properties and extend product shelf life.
Q: Where is this OTR testing equipment typically used?
A: This tabletop instrument is ideal for laboratories, quality assurance departments, and production facilities associated with edible oil packaging, particularly among manufacturers, suppliers, and exporters in markets requiring stringent packaging validation.
Q: What are the major benefits of using this OTR testing instrument for edible oil packaging?
A: Key advantages include compliance with global standards, accurate and repeatable measurements, automated operation, flexible sample handling (1-3 specimens), comprehensive data management, and immediate alerts for over-limit readings, directly supporting product quality and regulatory adherence.
Q: How does the alarm system function during OTR testing?
A: The system features both audible and visual alarms, providing real-time notifications if oxygen transmission exceeds set thresholds. This ensures prompt corrective actions, safeguarding both the test integrity and product quality.
Q: What software capabilities and data output options are included with the instrument?
A: Advanced analysis software allows multi-user access, data logging, and auto-calibration, while test results are easily exported in CSV and PDF formats for record keeping or further evaluation.