Product Description
High Precision
- Patented test chamber and test dishes with advanced fluid dynamics and thermodynamic designs.
- Labthinks proprietary thermostat technology ensures that the test chamber is precisely temperature controlled and remains stable throughout the test.
- Precise and scientific regulation and calculation of testing conditions.
High Efficiency
- 12 test stations
- Supports test modes of desiccant method and water method
Labor Saving
- Automatic humidity and flow rate controlling.
- Desiccator requires no replacement of inner core.
- High-efficiency water vapor generator.
Simplified Operation
- 12 touch-screen pad powered by WindowsTM 10 operating system
- Fast automatic testing process
- Optional DataShieldTM software and accessories for automatic data management
Product Features
New Generation Test Chamber and Test Dishes
Patented test chamber and test dishes with advanced hydrodynamic and thermodynamic designs ensure the uniform flow rate over the specimen surface, stable temperature and humidity, creating a uniform and stable testing environment. As a result, the test duration is shortened and the test results will be more accurate.
Excellent Testing Abilities of High and Low Barrier Materials
The testing conditions will be regulated precisely in real time, which ensures the high testing accuracy and repeatability of high and low barrier materials.
Automatic Controlling of Temperature, Humidity and Air Velocity
Labthinks proprietary thermostat technology ensures that the test chamber is precisely temperature controlled and remains stable throughout the test.
Air velocity is monitored and regulated in real time.
High-efficiency and mist-free automatic humidity regulator meets the requirements for long-term continuous tests.
It is unnecessary to replace the inner core of the desiccator, which can continuously work for 20,000 hours.
Easy-to-use and High-efficiency System
The automatic test mode, combined with the instrument features, eliminates the need for manual adjustments to quickly obtain accurate results, saving training costs and releasing staff from manual monitoring so that they are available for other tasks.
The professional test mode provides flexible and rich instrument control functions to meet individual scientific research needs.
Unique, optional DataShieldTM system facilitates centralized management of user data. It supports a variety of formats of exported data. Reliable security algorithms are used to prevent data leakage. It supports universal wired and wireless LAN, optional private wireless network and third-party software.
User-oriented Service Concept
Adhering to our user-oriented service concept, Labthink has created a customization system that provides flexible and comprehensive customization services for the accommodation of non-standard specimens and packages.
Test Principle
The test specimen is mounted in the test dish which contains water or desiccant inside. The test dish is placed in the test chamber with stable temperature, humidity and air flow. The water vapor permeates through the specimen and into the dry side. By measuring the weight changes of the test dish periodically, water vapor transmission rate and other parameters can be obtained.
Test Standards
ASTM E96, ASTM D1653, ISO 2528, TAPPI T464, DIN 53122-1, GB/T 1037, GB/T 16928, YBB 00092003-2015
Applications
| Basic Applications |
- Films
Plastic films, plastic composite films, paper-plastic composite films, coextruded films, aluminum coated films, aluminum foil composite films, glass fiber aluminum foil paper composite films and many other film materials
|
- Sheeting
PP, PVC and PVDC sheeting, metal foils, rubber pads and other sheeting materials
|
- Paper and Paper Board
Aluminum coated paper for cigarette, paper aluminum plastic composite film and other paper and paperboards
|
- Textiles and Nonwovens
Textiles and nonwoven fabrics
|
| Extended Applications |
- Construction Materials
Geotextiles, felt, roofing and building materials, vapor barrier membranes, etc.
|
- Aseptic Wound Protection Films and Medical Plasters
Aseptic wound protecting films, medical plasters and protective clothing materials
|
Technical Specifications
Table 1: Test Parameters Note1
| Parameter |
Model C360H |
| Testing Efficiency |
0.01g/(m2day) 0.5g/(m2day) |
>24 hours |
| 0.0006g/(100in2day) 0.0323g/(100in2day) |
| 0.5 g/(m2day)5 g/(m2day) |
12 24 hours |
| 0.0323g/(100in2day) 0.3225g/(100in2day) |
| > 5 g/(m2day) |
12 hours |
| > 0.3225 g/(100in2day) |
| Max. Test Range |
Water Method |
10000/ni141-12i14g/(m2day) |
| 645/ni141-12i14g/(100in2day) |
| Desiccant Method |
1200g/(m2day) per piece |
| 77g/(100in2day) per piece |
| Test Station |
|
12 |
| Test Temperature |
C |
20 55 0.2 |
| Test Humidity |
RH |
10% 90% 1% |
| Additional Functions |
DataShieldTM Note2 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Specimen Size |
I74mm |
| Specimen Thickness |
3mm |
| Test Method |
Desiccant Method / Water Method |
| Standard Test Area |
33cm2 |
| Carrier Gas |
Compressed Air |
| Drying of Carrier Gas |
Long Service-life Desiccator (unnecessary to replace inner core) |
| Humidifying of Carrier Gas |
High-efficiency mist-free humidity generator |
| Carrier Gas Pressure |
0.6 MPa |
| Port Size |
I6mm PU Tubing |
Note 1: The parameters in the table are measured by professional operator in Labthink laboratory according to relative requirements for laboratory standard conditions.
Note 2: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be purchase as required.
Note 3: The described product features should be in line with Table 1: Test Parameters.
Precise Environmental Control for Reliable ResultsThe Automated WVTR Tester features an integrated microclimate enclosure, ensuring meticulous regulation of temperature (15C-55C) and humidity (20%-90% RH). This environmental control is vital for consistent, reproducible measurements, particularly in the sensitive evaluation of medical hydrogel dressings where WVTR performance directly impacts patient outcomes. Programmable settings make it easy to simulate real-life storage or application conditions.
User-Friendly Operation and Flexible Data ManagementOperating the tester is streamlined by an intuitive Windows-compatible software interface, touch-screen controls, and quick-lock leak-proof sample clamps. Users can log and export up to 9999 test records in CSV or PDF format, facilitating smooth data management and compliance documentation. Automated and manual calibration options minimize downtime and reduce training requirements.
Comprehensive Support for Quality AssuranceTo ensure rapid deployment and ongoing reliability, the manufacturer offers onsite installation and remote online diagnostics. Built-in alarms for over-temperature and humidity conditions increase operational safety. These features, combined with full compliance to international standards, make this tester ideal for quality control and R&D applications in the production of medical hydrogel dressings.
FAQ's of The Benefits of WVTR Testing for Medical Hydrogel Dressings:
Q: How does WVTR testing benefit manufacturers of medical hydrogel dressings?
A: WVTR (Water Vapor Transmission Rate) testing allows manufacturers to evaluate the moisture management properties of hydrogel dressings, ensuring optimal breathability and fluid control. By verifying compliance with ASTM E96 and ISO 2528, producers can guarantee consistent product performance, reduce the risk of wound maceration, and meet regulations in global wound care markets.
Q: What is the process for conducting a WVTR test using this equipment?
A: After installing specimens (100 mm 100 mm) in the quick-lock clamps, users select desired test parameters through the digital LCD touch interface. The integrated microclimate enclosure maintains specified temperature and humidity. Tests are performed using gravimetric or sensor-based detection, and the instrument automatically logs and stores data, which can be exported in CSV or PDF formats for analysis.
Q: When should calibration of the WVTR tester be performed?
A: Calibration should be performed upon initial installation, after major maintenance, or as part of routine quality assurance protocols. The tester supports both automatic and manual calibration options, ensuring measurement accuracy and compliance with regulatory standards over time.
Q: Where can this WVTR tester be installed, and what are the mounting requirements?
A: The Automated WVTR Tester is designed as a benchtop unit suitable for laboratories, manufacturing quality control rooms, or R&D facilities. It requires a stable surface, operating voltage of 220V AC at 50 Hz, and standard lab conditions with appropriate gas pressure and environmental controls.
Q: What are the key features that enhance workflow efficiency and safety during WVTR testing?
A: Efficiency is boosted by features such as quick-lock sample clamps, fully automatic data logging, and programmable test parameters. Operational safety is ensured by real-time over-temperature and humidity alarms, an enclosed testing environment, and onsite or remote service support.
Q: How is test data managed and reported from this device?
A: Test results-including up to 9999 records-are automatically stored on the device and can be exported as CSV or PDF reports via USB or RS232 interface. This flexible data management supports audit trails, traceability, and regulatory compliance.