Product Description
Product FeaturesNote1
- Handheld and portable design, easy-to-operate with one hand, suitable for tests on production line
- Auto shutdown for energy saving
- One-button calibration, simple and efficient
- Components and parts are supplied by global well-known manufacturers, which guarantee stable performance of the instrument
- Protective cover for sampling probe to ensure the safety of the operator
- Embedded gas sensors for accurate analysis of the gas contents inside flexible or rigid packages
- Embedded pressure sensor for accurate determination of the vacuum degree inside the package
- CO2 sensor uses NDIR technology, patented solid state LED, detector and gold-plated optic components
- The sensors are supplied by globally well-known manufacturers, which have ultra-high precision, low fault rate and long service life
- Industrial grade touch screen, one button operation, intuitive operating interface, remote upgrade & maintenance
- Chinese and English language operating system can meet different requirements of the users
- Automatic data storage, power failure memory to prevent data loss
- Up to 1200 test results can be stored
- Equipped with wireless micro-printer for convenient data printing (optional)
- Equipped with standard USB ports and professional software for convenient PC connection and data transfer (optional)
Test Principle
The gas inside the test package is brought to the gas sensor by the air pump. The instrument analyzes the volume of oxygen and carbon dioxide (optional) contained in the sample gas by calculating the electrical signals generated by the gas sensor. When the preset condition is met, the test is completed and the system records the oxygen and carbon dioxide (optional) concentration.
Applications
| Basic Applications |
- Packaging Bags
Test the volume of O2 and CO2 (optional) in headspace of non-vacuum sealed packaging bags of coffee, cheese, milk tea, milk powder, bread, bean powder, instant food and drugs, etc.
|
- Packaging Containers
Test the volume of O2 and CO2 (optional) in headspace of packaging containers of coffee, milk powder, food, cheese, can, Tetra Pak and beverage, etc.
|
| Extended Applications |
- Ampoule Bottles
Test the volume of O2 and CO2 (optional) in headspace of ampoule bottles
|
Technical SpecificationsNote2
| Specifications |
C650M |
| Testable Gases |
O2 (Standard) |
CO2 (optional) |
| Test Principle |
Electrochemistry |
Infrared Absorption |
| Sensor Service Life |
2 Years (in air) |
>15 Years |
| Test Range |
0 100% |
0 100% |
| Resolution |
0.01% |
0.01% |
| Test Accuracy |
0.2% |
(0.03% + displayed value * 5%) |
| Sampling Volume |
<2 mL (Standard Mode) |
15 mL (Standard Mode) |
| Instrument Dimension |
220mm (L) 110mm(W) 70mm(H) |
| Power Supply |
220VAC10% 50Hz / 120VAC10% 60Hz |
| Net Weight |
0.6 kg |
Configurations
- Standard Configurations
- Instrument, Sampling Probe, Filter, Seal Gasket
- Optional Parts
- Professional Software, Micro Printer, B2227 Test Frame for Headspace Gas Analyzer, B2226 Underwater Sampling Set, CO2 Sensor, Pressure Sensor
Note 1: The described test standard, applications and product features should be in line with Technical Specifications.
Note 2: The parameters in the table are measured by professional operators in Labthink laboratory under strictly controlled laboratory conditions.
Note 3: The picture in this brochure has included B2227 Test Frame for Headspace Gas Analyzer.
Enhance Drug Safety and Quality ControlDetecting residual oxygen in pharmaceutical packaging ensures product efficacy, shelf life, and regulatory compliance. The C650M Headspace Gas Analyzer delivers reliable and precise oxygen, nitrogen, and air analysis for vials and ampoules, supporting manufacturers in meeting strict safety standards while protecting product integrity.
Productivity Meets ComplianceWith a capacity of up to 60 samples per hour and data memory for 10,000 records, the analyzer streamlines high-volume testing environments. Its compliance with GMP and 21CFR Part 11 regulations assures pharmaceutical exporters and suppliers of robust, audit-ready documentation. Microcomputer control and automated calibration facilitate consistent, error-free results.
Versatile and User-Friendly DesignThe C650M is engineered for bench-top use with adjustable test speed and customizable settings, suitable for varied specimen sizes and types. Its intuitive digital touch screen allows for straightforward operation in multiple languages, simplifying workflows for teams worldwide. The device's non-destructive testing approach preserves valuable samples for further analysis.
FAQ's of Improve Drug Safety with C650M Headspace Gas Analyzer:
Q: How does the C650M Headspace Gas Analyzer improve drug safety in pharmaceutical packaging?
A: The analyzer provides precise detection of residual oxygen within pharmaceutical vials and ampoules, minimizing the risk of oxidation that could compromise drug potency and shelf life. Reliable measurements support compliance with strict safety standards, ensuring product quality throughout the supply chain.
Q: What types of gases can the C650M test for, and which packaging formats are compatible?
A: The C650M is compatible with oxygen, nitrogen, and air analysis. It is specifically designed for headspace gas testing in single vials or ampoules ranging from 10 to 100 mm in height, making it ideal for pharmaceutical packaging applications.
Q: When should automatic or manual calibration be performed on the analyzer?
A: Automatic calibration is recommended for routine quality assurance and minimizing operator intervention, while manual calibration is available for protocol-specific requirements or when revalidating instrument accuracy after significant maintenance or environmental changes.
Q: Where can I access and store measurement data from the analyzer?
A: Measurement data can be stored on the device's integrated memory, which accommodates up to 10,000 records. For enhanced documentation and printing, data can be exported via RS232/USB connections, and printed using an optional external printer, suitable for GMP-compliant record retention.
Q: What is the process for testing sample packaging with the C650M?
A: Place a single vial or ampoule specimen in the analyzer, select measurement parameters via the touch screen interface, and initiate testing. The device provides results within ten seconds, with adjustable test speed and customizable test width to suit sample dimensions, while the non-destructive method preserves the sample.
Q: How does the integrated alarm system and compliance features benefit operations?
A: The built-in over-limit alarm provides immediate notification of critical gas concentration levels, helping ensure user safety and product integrity. Compliance with GMP and 21CFR Part 11 enhances audit readiness and supports the establishment of standardized quality procedures.