Product Description
Product FeaturesNote1
Infrared sensor
- Equipped with Labthink patented infrared sensor, features a wider test range.
- Designed conforming to ASTM F1249.
- Ultra-long service life, non-consumable type.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of velocity, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click startup, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthinks unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with various intelligent sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
Standard
ASTM F1249ISO 15106-2GB/T 26253JIS K7129YBB00092003-2015
Applications
| Applications |
- Films
Water vapor transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others.
|
- Sheets
Water vapor transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C306H |
| Test range |
g/(m2day) (Standard area 50cm2) |
0.0240 |
| g/(m2day) (MASK area 5cm2) |
0.2400Optional |
| g/(m2day) (MASK area 1cm2) |
12000Optional |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.02 or 2%take the greater |
| Test Temperature |
|
1550560Optional |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHWithin standard temperature range |
100%5902% |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.611.7cm11.7cm |
| Specimen Thickness |
120 Mil3mm |
| Standard test area |
50cm2 |
| Carrier Gas |
99.999% High purity nitrogen(Outside of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzone of two |
| Net Weight |
220Lbs100kg |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, 6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is 6 mm PU tubingpressure79.7 PSI / 550 kPa, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Advanced MVTR Testing for Medical ApplicationsThis instrument is specifically engineered to deliver optimal MVTR testing for wound healing dressings and medical films. By adhering to ASTM E96 and ISO 12572 standards, it assures regulatory compliance and accurate data, critical for medical device innovation and approval.
User-Friendly, Programmable, and VersatileThe device offers automatic or manual control modes, a digital LCD for real-time display, and user-programmable calibration. Routine calibration and protective alarms safeguard reliable measurements while its semi-automatic design accommodates various sample types. Built-in preparation tools and adjustable environment controls simplify usage for laboratory professionals.
FAQs of Ensuring Optimal Wound Healing Dressings Through Accurate MVTR Testing Instrument:
Q: How does the MVTR Testing Instrument ensure accurate measurement of wound healing dressing performance?
A: The device utilizes user-programmable, factory calibration and advanced sensors to deliver highly precise MVTR results (accuracy of 0.01 g/m24h). By strictly following ASTM E96 and ISO 12572 standards, the instrument ensures each measurement reflects realistic clinical conditions.
Q: What is the process for preparing and testing samples with this instrument?
A: Samples are prepared using included tools and placed in anodized aluminum holders (74 mm diameter). The operator selects preferred control mode (automatic or manual), sets environmental parameters, and initiates the test. Real-time data is output digitally, and results can be exported in CSV or PDF format.
Q: When should the calibration of the instrument be performed?
A: Calibration can be performed as needed, either during installation or routine quality checks. The system supports user-programmable calibration and is factory calibrated, ensuring sustained measurement reliability for critical medical quality assurance.
Q: Where is this testing instrument best suited for use?
A: This benchtop instrument is designed for laboratory settings, particularly in medical device manufacturing, research institutions, and quality control labs. Its durability and compliance make it valuable for wound care product developers and medical film producers.
Q: What are the compliance standards supported by this MVTR instrument?
A: The instrument adheres to ASTM E96 and ISO 12572 standards, ensuring results are suitable for regulatory submissions and medical-grade quality control.
Q: How is test data output and accessed from the device?
A: Data is displayed in real-time on a digital LCD screen and can be exported via USB or RS232 interface in both CSV and PDF formats for easy documentation and analysis.
Q: What are the key benefits of using this MVTR Testing Instrument for wound healing dressings?
A: The instrument delivers rapid, highly accurate MVTR measurements, capacities for multiple specimens, and supports full regulatory compliance. Its programmable calibration, protective alarms, and comprehensive data outputs make it ideal for efficient product development, testing, and quality control.