Product Description
Product FeaturesNote1
Infrared Moisture Sensor
- The Labthink patented technology infrared moisture sensor can obtain lower test limit.
- Designed according to ASTM F1249.
- Long service life, non-consumable.
- It has the functions of over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor for real-time monitoring and recording temperature and humidity changes.
- In the test process, the automatic control of flow, temperature and relative humidity is realized, and the accuracy is higher.
- It can achieve higher test repeatability of 0.005 g / (m2day) .
High Efficient Three-chamber
- Independent three sets of 50cm2 standard area test chamber, in line with the standard requirements of parallel sample detection.
- Three samples are tested at the same time under the same condition, and the data are independent of each other.
- In the same test cycle, the testing number of samples completed increased from 2 to 3.
- Automatic sample clamping, saving time and effort, realizing consistent clamping force and better sealing.
Intelligent Control
- The 12 Inch Touch panel of Windows system is used to control, which is more convenient.
- Automatic mode: input the test temperature and humidity, click one key to start the automatic test.
- New drawer type test chamber, one key automatic in and out, sound and light reminder.
Safe and Reliable
- Safe running: the unique high-end industrial computer of Labthink is embedded to eliminate the system failure caused by computer virus, and ensure the operation reliability and data storage security.
- Safe operation: equipped with optical and other intelligent sensors, sound and light alarm and other intelligent alarms to ensure the safety of operation.
- Safe Performance: the instruments are made of world famous brand components, stable and reliable.
Powerful Functions
- Professional test mode provides flexible and rich control functions to meet the needs for scientific research.
- Provide water vapor transmission curve, water vapor transmission coefficient curve, temperature curve, humidity curve.
- Ultra wide test range, meet the barrier test of various materials. (Customized)
- Ultra wide temperature range to meet the permeability tests at different temperatures. (Customized)
Testing Principle
The pre-treated sample is clamped between the test chambers, the nitrogen with stable relative humidity flows on one side of the film, and the dry nitrogen flows on the other side of the film; due to the existence of humidity difference, water vapor will diffuse from the high humidity side through the film to the low humidity side; on the low humidity side, the water vapor is carried to the infrared water sensor by the flowing dry nitrogen, and different water vapor is generated. The concentration produces different light signals, and the concentration value is obtained through analysis and calculation, and then the water vapor transmission rate of the sample is calculated.
Reference Standards
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The water vapor transmission rate tests of various plastic film, paper plastic composite film, co-extrusion film, aluminum plating film, aluminum foil composite film, glass fiber aluminum foil paper composite film and other film materials.
|
- Sheet
The water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Parameters
Table 1: Test Parameters Note2
| Parameters/Model |
C303H |
| Test Scope |
g/(m2day) (standard area 50cm2) |
0.005i40 (standard) 0.005100i14optionali14 |
| g/(m2day) (MASK area 5cm2) |
0.05400 (optional) |
| g/(m2day) (MASK area 1cm2) |
0.252000 (optional) |
| Resolution |
g/(m2day) |
0.0001 |
| Repeatability |
g/(m2day) |
0.005 or2%, whichever is greater |
| Temperature Range |
|
15i50 560 (customized) |
| Temperature Fluctuation |
|
0.05 |
| Humidity Range |
%Rh (within standard temperature range) |
5i901% |
| Extended Functions |
DataShieldTM Data Shield Note 3 |
Optional |
| GMP Computer System Requirements |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 sets |
| Sample Size |
4.4 x 4.4i1411.2cm11.2cmi14 |
| Sample Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50cm2 |
| Carrier Gas Specification |
99.999% high purity nitrogen (gas source is prepared by the user) |
| Air Source Pressure |
40.6 PSI / 280 kPa |
| Interface Size |
1 / 8 "metal tube |
| Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Power Supply |
120VAC 10% 60Hz / 220VAC 10% 50Hz |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Mainframe, tablet computer, sampler, vacuum grease, polyurethane pipe with diameter of 6 mm |
| Options |
Air compressor, CFR21Part11, GMP computer system requirements, DataShieldTM Data Shield Notes3 |
| Note |
The compressed air inlet of the machine is I 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa); the air source is prepared by the user. |
Note 1: the product features are subject to the specific notes in the "Technical Parameters" table.
Note 2: the parameters in the table are measured by professional operators in Labthink laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield system provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase it separately according to the usage.
High-Precision Water Vapor AnalysisThe C303H employs advanced infrared detection and an electrolytic sensor to ensure accurate and stable water vapor transmission rate measurements. Achieving high repeatability ( 10%), this system is designed to meet the demanding standards of film, packaging, and textile testing while supporting robust automation and easy calibration.
Convenient & Versatile OperationOffering both touch screen and PC software control, the C303H integrates seamlessly into laboratory or industrial workflows. Its flexible chamber accommodates up to three specimens and supports various data output formats, such as CSV, PDF, and direct printing, maximizing lab efficiency and ensuring precise data management.
Comprehensive Data ManagementWith compatibility for the LystemTM LabData System, storing and retrieving up to 10,000 test records is streamlined. Real-time data updates, programmable test parameters, and auto data storage underpin enhanced traceability and regulatory compliance for critical testing applications.
FAQ's of C303H Water Vapor Transmission Rate Test System:
Q: How does the C303H measure water vapor transmission rate?
A: The C303H utilizes an advanced infrared detection method, with an optional gravimetric measurement. Water vapor permeating through the sample is detected by an electrolytic sensor, ensuring precise and immediate real-time data collection.
Q: What types of materials can be tested with the C303H system?
A: This system is designed to test a wide variety of materials, such as films, packaging materials, sheets, and textiles, making it suitable for quality control and research in numerous industries.
Q: When do I need to calibrate the C303H, and is the process automatic?
A: Calibration is essential for maintaining measurement accuracy and can be carried out automatically or manually, depending on your laboratory's standard operating procedures or specific quality requirements.
Q: Where can the C303H be installed, and what are the operating conditions?
A: The C303H is designed for benchtop use in both laboratory and industrial environments. It operates optimally in temperatures between 15-35C and relative humidity up to 80% RH.
Q: What is the process for conducting a test using the C303H?
A: Samples are securely clamped using the sealing ring and clamp mechanism. Test parameters, including humidity and temperature, can be programmed via the touch screen or PC software. After initiating a test, real-time results are displayed and saved automatically.
Q: How can I access and use the test data from the C303H?
A: Results can be exported directly in CSV or PDF formats, printed, or managed through the LystemTM LabData System. Up to 10,000 test results can be stored in memory for convenient retrieval and analysis.
Q: What are the key benefits of using the C303H Water Vapor Transmission Rate Test System?
A: The C303H offers high accuracy, programmable and automated functionality, multi-specimen testing, secure data management, and user-friendly operation, improving productivity and compliance in laboratory and industrial settings.