Product Description
Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cm) |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14either one of twoi14 |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbsi14100kgi14 |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument isI 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Reliable Barrier Performance AnalysisSpecially designed for milk powder packaging, this instrument provides accurate assessment of oxygen and water vapor barrier properties. Its compliance with international testing standards ensures results are suitable for quality assurance and research applications. The ability to analyze up to three specimens at once enables efficient and comparative testing workflows.
Intuitive User ExperienceWith an LCD touch screen and dual-language menu (English/Chinese), users can easily operate the device and access real-time data. The microcomputer control system automates the test process, while adjustable settings for speed, width, and stroke offer versatility to match diverse specimen requirements.
Comprehensive Data Management and SafetyThis instrument's robust built-in storage securely archives over 500 test records. Automatic data processing and curve display support detailed analysis, while overpressure automatic protection enhances operational safety. Multiple calibration options and included sample holders simplify routine maintenance and setup.
FAQ's of Instrument Measuring Milk Powder Packaging Barrier Performance:
Q: How does this instrument measure the barrier performance of milk powder packaging?
A: The instrument utilizes test gases (oxygen and water vapor) in a controlled stainless steel chamber to assess how much gas permeates through the packaging material. Sensors detect transmission rates in real-time, providing precision data on barrier effectiveness according to ASTM D3985 and ISO 15106-2 standards.
Q: What are the typical applications for this testing equipment?
A: Mainly used in research and quality control labs, the instrument evaluates packaging barrier properties to ensure milk powder remains protected from moisture and oxygen. This is critical for maintaining product shelf life and hygiene standards in the food packaging industry.
Q: When should I perform calibration on the instrument?
A: Calibration should be carried out before initial use, periodically as routine maintenance, or whenever measurement accuracy needs to be verified. The device supports both automatic and manual calibration, making it convenient to maintain optimal performance.
Q: Where can this testing device be installed and used?
A: This laboratory instrument is designed for indoor environments with non-vibrating surfaces. It requires a stable benchtop, standard 220V power, and suitable environmental conditions (humidity 20-90% RH, temperature 15-50C) for accurate and safe operation.
Q: What is the testing process for packaging barrier performance?
A: Users place up to three specimens in the sample holders, set testing parameters via the touch screen, and initiate the automated process. The instrument then runs the test within a 0.6 MPa gas pressure environment, records result data, and displays real-time curves for analysis.
Q: How can I benefit from using this instrument in my packaging operations?
A: Using this device ensures precise, quick, and reliable assessment of packaging barrier performance, helping optimize packaging materials and processes, improve product shelf life, and ensure compliance with international quality standards.
Q: What features make data handling and reporting efficient with this device?
A: The built-in data storage accommodates up to 500 test results, while automatic data processing and real-time LCD display ensure easy access and interpretation. Results can be exported through RS232 or USB interfaces for further analysis or documentation.