Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR \ WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR \ WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.411.2cm11.2cm) |
| Specimen Thickness |
120 Mil3mm |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzeither one of two |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbs100kg |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument is 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Precision Testing for ComplianceThis MVTR & OTR tester meets key international standards, including ASTM F1249, ASTM D3985, and ISO 15106-3. It delivers accurate and repeatable barrier evaluations, ensuring your seasoning blend packaging materials meet regulatory and quality benchmarks. Certified calibration options, programmable pre-conditioning, and automated processes enhance the reliability and efficiency of every test.
Customized Testing EnvironmentsUsers can program the system for customizable humidity (10%90% RH) and temperature (1550C) conditions to match specific client needs. Whether testing in a laboratory or production environment, its microprocessor-based control and wide dynamic range ensure precise results for a variety of sample materials. High-throughput capability enables simultaneous analysis, improving productivity.
User-Friendly Automation and Data ManagementThe system features a multilingual interface, automated test sequences, password-protected user modes, and graphical LCD display for real-time monitoring. Data can be stored internally or exported for further analysis using the included PC-based software. Alarm systems alert users to over-range or under-range results, safeguarding your testing process.
FAQs of MVTR & OTR Tests for Seasoning Blend Packaging Materials:
Q: How do I conduct MVTR and OTR tests for seasoning blend packaging using this system?
A: Begin by conditioning your specimens with the programmable pre-conditioning function. Load up to three 100 mm diameter samples into the chambers. Select the desired test parameters, such as humidity, temperature, and test speed via the microprocessor-based interface. The system performs automated testing and provides real-time, accurate results on the graphical LCD display.
Q: What compliance standards does the instrument fulfill for barrier property testing?
A: The system is compliant with ASTM F1249 for MVTR, ASTM D3985 for OTR, and ISO 15106-3. These standards ensure reliable, globally recognized testing protocols, making the instrument suitable for regulatory and quality assurance in food packaging.
Q: Where can the test results be stored and accessed?
A: Test data can be automatically saved internally within the instruments memory and exported via USB or RS232 interfaces. You can also manage and analyze data externally with the included PC-based analysis software, ensuring robust documentation and easy reporting.
Q: What is the process for calibrating the test system?
A: Calibration can be performed either at the factory or by trained users on-site. The system supports user-friendly calibration steps to maintain accuracy (1% of reading), ensuring long-term reliable operation and consistent performance.
Q: When should sample conditioning be used before testing?
A: Sample conditioning is recommended to ensure reproducibility and reduce data variation caused by environmental differences. The programmable pre-conditioning module lets you set the appropriate duration, temperature, and humidity for each sample before testing begins.
Q: How does the built-in alarm system enhance safety during testing?
A: The instruments alarm system monitors all test parameters and instantly notifies users if readings exceed pre-set over-range or under-range limits. This feature helps protect both the samples and the instrument, ensuring test integrity and user safety.
Q: What are the main benefits of using this instrument for packaging barrier analysis?
A: Key benefits include automated, high-throughput barrier testing, compliance with major international standards, customizable test conditions, multilingual operation, secure user access, and detailed data management. These features streamline your workflow while delivering precise, reliable results for your seasoning blend packaging materials.