Product Description
Product FeaturesNote1
Innovative structure design & upgraded automatic operation
- Test cells with brand new pop-up drawer type design
- Automatic specimen clamping with uniform force
- Components are supplied by global well-known manufacturers, which are reliable and stable in performance
- Intelligent test mode, one-button test operation (test temperature should be set in advance) and automatic stop
- Scientific research mode provides more flexible parameter and function settings for analysis of gas transmission rate, solubility coefficient, diffusion coefficient and permeability coefficient
Upgraded testing accuracy and efficiency
- Imported high precision vacuum sensor for tests of ultra-high barrier property (0.01 0.09 cm3/ m224h0.1MPa) materials with high accuracy and repeatability
- Imported pneumatic control system with ultra-low failure rate and extremely long service life, which guarantees the seal performance of the whole system
- Imported high precision vacuum pump improves the evacuation efficiency and generates vacuum up to 0.2Pa
- The vacuum pump is turned on/off automatically
- For low and medium barrier property materials, test time is less than 4 hours (including evacuation time)
- For high barrier property materials, test time is less than 8 hours (including evacuation time)
- Three independent test cells provide individual test results, specimens can be replaced as desired
Excellent temperature and pressure control technique
- 360circulation constant temperature technology is applied in the instrument to remain the test temperature fluctuation below 0.05C
- Labthinks unique pressure compensation technique is used to maintain the stable differential pressure (pressure change is less than 0.2KPa)
- The pressure of high pressure chamber can be adjusted from 10KPa to 210KPa and maintained precisely
Versatile functions based on standards, customization available
- Various types of gases are testable: sole gas, mixed gases, poisonous gases, explosive gases and other dangerous gases (customization is required)
- The humidity of test gas can be adjusted and controlled automatically by the system, without requiring human intervention (customization is required for humidifying device)
- Labthinks unique data fitting function can provide the test results of gas transmission rate, permeability coefficient, solubility coefficient and diffusion coefficient at extreme temperatures
- Reference film for fast calibration ensures accurate and universal test data
- Calibration port for verifying test temperature and pressure
- Conform to the test standards of differential pressure method
High end, secure and easy-to-use embedded computer controlled system
- Integrated design of instrument and software minimizes the malfunctions caused by computer virus or faulty operations and ensures the performance of the instrument and data security
- The instrument can be easily operated with a mouse, a keyboard and a monitor. Windows operation interface for testing operations and displaying data
- The system is equipped with four USB ports and dual Internet ports for convenient data transmission
- The instrument meets the requirements of China GMP of pharmaceutical industry (optional)
- Labthinks unique DataShieldTM system for data management and connecting information management system (optional)
Test Principle
C130H is designed in accordance with differential pressure method. The pre-conditioned specimen is mounted in the gas diffusion cell as to form a sealed barrier between two chambers. The lower-pressure chamber is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber and a constant pressure difference is generated between the two chambers. The gas permeates through the specimen from higher pressure side into the lower pressure side. The gas permeability and other barrier properties of the specimen can be obtained by monitoring the pressure changes in the lower chamber.
Test StandardNote1
ISO 2556, ISO 15105-1, GB/T 1038-2000, ASTM D1434, JIS K7126-1, YBB 00082003
ApplicationsNote1
This instrument is applicable to the determination of gas permeability of:
| Basic Applications |
- Films
Including plastic films, plastic composite films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films and many others
|
- Sheeting
Including engineering plastics, rubber and building materials, e.g. PP, PVC and PVDC
|
| Extended Applications |
- Various Gases
Test the permeability of various types of gases, e.g. O2, CO2, N2, Air and He
|
- Inflammable, Explosive and Poisonous Gases
Test the permeability of inflammable, explosive and poisonous gases
|
- Biodegradable Films
Test gas permeability of various sorts of biodegradable films, e.g. starch-based biodegradable bags
|
- Materials for Aerospace Usage
This instrument can test the Helium permeability of airship gas bags
|
- Paper and Paper Board
Test gas permeability of paper and paper-plastic composite materials, e.g. aluminized paper for cigarette packages, Tetra Pak sheeting, paper bowls for instant noodles and disposable paper cups
|
- Paint Films
Test gas permeability of substrates coated paint films
|
- Glass Fiber Cloth and Paper
Including glass fiber cloth and paper materials, e.g. Teflon paint cloth, Teflon welding cloth and Teflon silicon rubber cloth
|
- Soft Tube Materials for Cosmetics
Including various types of cosmetic tubes, aluminum-plastic tubes and toothpaste tubes
|
- Rubber Sheeting
Including various sorts of rubber sheeting, e.g. car tires
|
Technical SpecificationsNote2
| Specifications |
Film Test |
| Test Range |
0.01 50,000 cm3/m224h0.1MPa |
| Resolution |
0.001 cm3/ m224h0.1MPa |
| Temperature Range |
10C 50C (room temperature 23C) |
| Resolution |
0.01C |
| Temperature Fluctuation |
0.05C |
| Temperature Accuracy |
0.3C (calibration port) |
| Vacuum Resolution |
0.01Pa |
| Vacuum Accuracy |
Displayed Value 0.2% (1%100% of sensor range) |
| Vacuum Degree of Test Chamber |
< 10 Pa |
| Number of Specimens |
3 with independent test results Customization available for other number of specimens |
| Specimen Size |
97 mm |
| Test Area |
38.48 cm2 |
| Test Gas |
O2, N2, and CO2 (outside of supply scope) Humidifying of Test Gas (customization available) |
| Test Pressure |
10KPa 210KPa |
| Gas Supply Pressure |
0.5 Mpa 0.6 Mpa (73psi 87psi ) |
| Port Size |
6 mm PU Tubing |
| Instrument Dimension |
710 mm (L) 350 mm (W) 630 mm (H) |
| Power Supply |
220VAC10% 50Hz / 120VAC10% 60Hz |
| Net Weight |
100 kg |
Configurations
- Standard Configurations
- Instrument, Monitor, Keyboard, Mouse, Round Sample Cutter, Vacuum Grease, Fast Quantitative Filter Paper and Vacuum Pump, 6 mm PU tubing (3m)
- Optional Parts
- GMP Computer System, DataShieldTM Note3, Air Compressor
- Note
- 1. The gas supply port of the instrument is 6 mm PU tubing;
2. Customers need to prepare gas supply.
Note 1: The described test standard, applications and product features should be in line with Technical Specifications.
Note 2: The parameters in the table are measured by professional operators in Labthink laboratory under strictly controlled laboratory conditions.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Advanced Equal Pressure Testing TechnologyThis analyzer harnesses the equal pressure differential method to accurately measure the gas transmission rate of film and packaging materials. The automated calibration, adjustable test parameters, and option for various carrier gases ensure reliable results for diverse industry needs. With a 7-inch color LCD touchscreen and multi-mode testing, operation is simple and efficient for every user.
Comprehensive Compliance and User-Focused DesignComplying with international standards such as ASTM D3985 and ISO 15105-2, the analyzer delivers precision and consistency. Its stainless steel chamber and integrated oil-free vacuum pump promote durability and low maintenance. User-friendly features like tool-free sample loading and automated data storage streamline workflows for laboratories and production environments.
Flexible Applications and ReportingDesigned to accommodate up to three specimens at once, the analyzer is ideal for film permeability and barrier property analysis. Real-time data display, a wide measuring range, and export capabilities mean test results can be easily reviewed and shared. The device fits seamlessly in research, quality control, and manufacturing settings.
FAQs of Analyzer for Film Gas Transmission Rate with Equal Pressure Method:
Q: How does the Analyzer for Film Gas Transmission Rate with Equal Pressure Method operate?
A: The analyzer uses the equal pressure differential method, where a test film divides two chambers with equal pressure. Selected carrier gases (oxygen, carbon dioxide, or nitrogen) pass through the film while sensors (electrochemical or zirconia) detect the permeated gas quantity. This setup allows precise, real-time monitoring of the films barrier properties.
Q: What types of films and applications is this analyzer suitable for?
A: This equipment is designed for use in film and packaging industries, supporting testing of plastic films, composite films, and barrier materials. Its applicable in quality control, research, and manufacturing where accurate gas permeability data is crucial for product performance and shelf life.
Q: When should I calibrate the analyzer, and what options are available?
A: Calibration is recommended at regular intervals or after critical maintenance. The instrument provides both automatic and manual calibration functions, ensuring sustained accuracy and compliance with ASTM and ISO standards.
Q: Where can the Film Gas Transmission Rate Analyzer be installed?
A: The analyzer is a benchtop unit suitable for laboratories or production facilities. It operates effectively in ambient conditions of 5C35C, with adjustable temperature and humidity controls. Ensure a stable surface and proper power supply for consistent performance.
Q: What is the process for loading and testing specimens?
A: Samples, up to three at a time, are loaded using a tool-free fixture that ensures quick and secure placement. Users select test parameters through the touchscreen interface, and the system conducts the test automatically, with adjustable speed and test duration according to sample and requirement.
Q: How does the analyzer manage and export test data?
A: Test results are automatically stored in the devices internal memory. Data can be exported via USB or RS232 interfaces and is compatible with Windows-based analytical software, allowing easy review, reporting, and record-keeping.
Q: What are the main benefits of using this analyzer in film barrier property analysis?
A: The analyzer offers high accuracy (1% F.S.), rapid response times, simultaneous multi-sample testing, and robust compliance with international standards. Its low-maintenance design, automated operation, and flexible data handling significantly improve laboratory efficiency and testing reliability.