Product Description
Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cm) |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14either one of twoi14 |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbsi14100kgi14 |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument isI 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Comprehensive Film Permeability AnalysisThe analyzer provides robust testing capabilities for both oxygen and water vapor transmission rates across films of varying thicknesses. By offering high accuracy and resolution, it allows users to quantify subtle variations in permeability, making it invaluable for research, quality control, and product development in packaging, materials science, and manufacturing sectors.
Efficient, Automated OperationAutomation is at the heart of this instrument. With microcomputer automatic control, multiple testing modes, and features such as automatic temperature and humidity adjustments, users benefit from reduced manual intervention and heightened consistency. The touchscreen interface and adjustable sampling intervals further enhance user experience, facilitating straightforward operation and data management.
Safe and Customizable Testing EnvironmentEquipped with over-temperature and leakage alarms, the analyzer ensures reliable and safe operation throughout testing. Its stainless steel test chamber and automatic calibration maintain measurement integrity, even during prolonged use. Adjustable test speed, temperature, humidity, and specimen number provide the flexibility to match specific application requirements.
FAQ's of Measuring the Impact of Film Thickness on Film Oxygen and Water Vapor Transmission:
Q: How does the analyzer measure the impact of film thickness on oxygen and water vapor transmission rates?
A: The equipment uses high purity nitrogen or oxygen to evaluate permeability, following ASTM D3985 and ASTM F1249 protocols. By allowing film samples up to 3 mm thick, it determines how thickness affects OTR and WVTR, with precise results stored automatically. Adjustable variables let users optimize measurements for different material types and thicknesses.
Q: What is the required process for operating this analyzer?
A: To begin, prepare your film specimen and place it in the stainless steel test chamber. Set desired temperature, humidity, and sampling interval, and select the appropriate testing mode via the 7-inch touchscreen. The microcomputer controls the testing automatically, and results can be exported through RS232 or USB interfaces.
Q: Where can this oxygen and water vapor transmission rate analyzer be typically used?
A: This instrument is ideal for laboratories, industrial facilities, and research institutes focused on material science, packaging development, or film production. Its robust construction and automation features support both small-scale testing and regular quality assurance procedures in varied environments.
Q: When should calibration be performed, and how is it managed?
A: Calibration is streamlined with the built-in reference film, and the analyzer automatically conducts self-calibration before measurement cycles. This guarantees consistent accuracy and reliability, recommended before each set of tests or when switching specimen types.
Q: What is the advantage of automatic temperature and humidity control during testing?
A: Automatic control ensures optimal testing conditions, stabilizing environmental variables that could impact OTR and WVTR measurement. This enhances reproducibility and accuracy across different tests, regardless of ambient fluctuations, which is particularly crucial for comparative studies or regulatory compliance.
Q: How are over-temperature and leakage incidents managed by the analyzer?
A: The analyzer is equipped with alarm systems that alert users to over-temperature or leakage situations immediately, prompting corrective actions and safeguarding both samples and operators during use.
Q: What benefits do adjustable data storage and sampling intervals offer for film testing?
A: With the capability to store up to 10,000 results and adjust sampling intervals from 1 to 60 minutes, users can efficiently manage multiple tests, conduct trend analysis, and ensure optimal resolution and reliability for both single specimens and batch studies.