TSY-W3 Water Vapor Permeation Test utilizes electrolytic detection method. It is a basic type and designed for WVTR test of flexible films, sheets and foils.
Principle:
Under a certain test temperature, a constant humidity difference is generated between two sides of the test specimen. The water vapor permeates through the specimen into the dry side and then is taken to the sensor, where proportional electric signals will be generated. The water vapor transmission rate and other parameters can be obtained by analyzing and calculating these electrical signals.
Standards:
Features:
Technical Specification
Specifications | Film Test |
Test Range | 0.001 ~ 50 g/m2·24h (standard) |
0.01 ~ 1000 g/m2·24h (optional) | |
Number of Specimens | 1 |
Resolution | 0.1ppm |
Temperature Range | 5°C ~ 95°C |
Temperature Accuracy | ±0.1°C |
Humidity Range | 0%RH, 2%RH ~ 98.5%RH, 100%RH (standard is 90%RH) |
Accuracy | ±1%RH |
Test Area | 38.48 cm2 |
Thickness | < 1 mm (accessories required for thicker specimens) |
Specimen Size | 100 mm |
Carrier Gas | 99.999% High Purity Nitrogen (outside of supply scope) |
Gas Flow | 100 mL/min |
Gas Supply Pressure | 0.12 MPas |
Port Size | 1/8 inch copper tubing |
Instrument Dimension | 500 mm (L) x 400 mm (W) x 360 mm (H) |
Power Supply | AC 220V 50Hz |
Net Weight | 36 kg |
Applications: