Product Description
Featuresnote1
Batch Testing
- Equipped with fully-automatic gravity fed sample feeder, which is Labthinks latest scientific and technological achievement that can facilitate continuous testing of 20 to 120 samples.
- A & B dual-channel design can test samples of two different specifications simultaneously.
- The target vacuum degree can be set by the operator to meet testing requirements of different samples.
- Dual methods of vacuum decay and pressure decay provide the flexibility to accommodate various types of package samples.
Precise Data
- Advanced pressure detection technology, using world renowned components for data stability which is not affected by ambient environment.
- Advanced microflow automatic flow control technology that can accurately simulate different sizes of leakage holes without manual adjustment.
- Both the differential pressure transducer and flow meter are traceable to NIST.
- The system can achieve a higher test repeatability of 1um.
Intelligent Control
- 15.6" embedded touch tablet computer with Windows OS.
- New longitudinal interface layout, and easy to use graphic user interface.
- Automatic alarm, automatic capture, automatic collection of unqualified samples.
- The pressure curve is displayed in real time, and the test results are counted automatically.
- Leakage rates are calculated automatically.
- The system is equipped with various sensors as intelligent reminders for safer operation and control.
- Universal printer can be connected for test results output.
- The system features embedded with USB and network ports to facilitate the external access and data transmission of the system, which can be upgraded remotely
Security Compliance
- Verified by compensation and calibration methods.
- The leak tester meets GMP requirements for data traceability and meets the needs of the pharmaceutical industry.
- User operation permission is managed at multiple levels, and the permission content can be configured on demand.
- Electronic signature is designed according to the standard requirements of 21 CFR Part11.
Test Principle
The sample is placed in the sample feeder and automatically delivered into the test cell. The leak rate and other results can be calculated and obtained by analyzing pressure changes measured by the sensor.
Reference Standards
ASTM F2338, YY-T 0681.18, and USP<1207>
Applications
| Basic Applications |
Vials |
Various vial sealing test. |
| Extended Applications |
Ampoule |
Various ampoule bottle sealing tests. |
| Cartridge Bottles |
Various cartridge bottle sealing tests. |
| Injection bottles |
Various injection bottles sealing test. |
Technical Parameters
Table 1: Test Parameters Note2
| ParameterModel |
C690H |
| Testing Range |
um(Reference aperture size USP1207) |
3i8igreat leakage |
| Detection Lower Limit |
um |
3 |
| Resolution |
um |
0.1 |
| Repeatability |
um |
1 |
| Pressure Range |
kPa |
-100i0ii14100 |
| Extended Functions |
21 CFR Part11 |
Optional |
| GMP computer system requirements |
Optional |
Table 2: Technical Specifications
| Testing Cell |
1 set for A group and 1 set for B group |
| Sample Feeder |
1 set for A group and 1 set for B group |
| Sample Size |
I45 mm80mmNote3 |
| Sample Quantity |
20120 pieces |
| Gas Specifications |
Compressed Air (Gas source is provided by the user) |
| Gas Source Pressure |
40.6 PSI / 500 kPa |
| Port Size |
I6 mm Polyurethane tube |
| Dimensions |
33.4 H x 19.6 W x 28.7 D (85cm 50cm 73cm) |
| Power |
120VAC10% 60Hz / 220VAC10% 50Hzi14select one from the twoi14 |
| Net weight |
209Lbs (95kg) |
Product Configuration
| Standard Configuration |
Mainframe, embedded tablet computer, software, flow meter, European vacuum pump, I6 mm polyurethane tube |
| Customization |
Test cell and sample feeder of group A, test cell and sample feeder of group B, negative standard reference sample and positive standard reference sample designed according to sample specifications |
| Optional Parts |
GMP computer system requirements, 21 CFR Part11, air compressor, IQ/OQ/PQ documents |
Note 1: The described product characteristics are subject to the specific annotation of the "Technical Parameters" table.
Note 2: The parameters in the table are measured in the Labthink laboratory by professional operators according to the requirements and conditions of the relevant laboratory environmental standards.
Note 3: Group C test cell can be customized for samples beyond the "Sample Size", but the lower detection limit and test range will change according to the sample size, and the actual delivery shall prevail.
Innovative Non-Destructive Leak TestingOur system employs pressure decay/vacuum leak non-destructive technology, enabling rapid and accurate detection of leaks in vacuum blood collection tubes. This ensures each tube's integrity is verified without causing damage. The adjustable test stroke and customizable test width make it ideal for varied tube sizes and types, supporting laboratories requiring stringent quality assurance.
Fully Automated Workflow for High ThroughputEquipped with PLC automatic control, the machine achieves an impressive throughput of up to 120 tubes per minute, handling multiple specimens in customizable trays. The automatic rejection system and real-time alarm features guarantee that only flawless tubes proceed, streamlining quality control while minimizing manual labor.
Comprehensive Data and Seamless ConnectivityThe leak detector supports Modbus and TCP/IP protocols for easy integration into laboratory information systems. Test records and statistics are easily exportable via USB for compliance, analysis, and traceability. Multi-language operation and an intuitive 7-inch touch screen enhance user experience for diverse teams.
FAQ's of Automated Non-Destructive Leak Detection for Vacuum Blood Collection Tubes:
Q: How does the pressure decay/vacuum leak detection method work in this system?
A: The system applies pressure decay and vacuum technology, monitoring for pressure changes indicating leaks. This non-destructive technique ensures each vacuum blood collection tube is tested without compromising its structure or sterility.
Q: What types and sizes of tubes can the machine test?
A: It supports PET and glass vacuum blood collection tubes in sizes 1375 mm, 13100 mm, and 16100 mm. The test width and stroke are customizable, making it suitable for a variety of tube types and capacities.
Q: When and where is this equipment ideally used?
A: This leak detector is ideal for laboratory or industrial quality control environments, particularly in facilities focused on medical supply manufacturing or blood collection tube production.
Q: What is the process for handling defective tubes during testing?
A: Tubes that fail detection are automatically identified and rejected from the production line. Visual and audible alarms promptly notify operators about faulty samples, ensuring prompt action and minimal contamination risk.
Q: How user-friendly is the equipment in terms of maintenance and operation?
A: Routine cleaning and maintenance can be performed tool-free. The 7-inch touch screen HMI offers easy navigation with multilingual options, supporting both English and Chinese, making operation straightforward even for new users.
Q: What are the major benefits of using this automated leak detector for vacuum blood collection tubes?
A: Major benefits include high-speed, non-destructive integrity testing, reduced human error, reliable automatic rejection of defects, comprehensive data storage, and seamless integration with laboratory information systems, leading to improved productivity and quality assurance.