Product Description
Product Featuresnote1
OTR/WVTR tested in One Instrument
- OTR WVTR mode, OTR mode and WVTR mode are available for selection.
- OTR and WVTR tests can be automatically completed for each sample mounting.
- Genuinely reflect the OTR WVTR performance of the same sample, avoiding effect on test data due to sample contamination arising from instrument switching.
Coulometric Oxygen Sensor
- The instrument is equipped with a ppb level Coulometric Oxygen Sensor, as one of new Labthink technical achievements, which can obtain a lower limit of the test.
- Designed according to ASTM D3985 with absolute value, and no calibration is permitted.
- Longer service life, two times longer than that of a traditional Coulometric oxygen sensor.
- Over limit alarm and automatic protection.
Infrared Moisture Sensor
- The instrument is equipped with Labthink patented infrared moisture sensor which can obtain a lower limit of the test.
- Designed according to ASTM F1249.
- Longer service life, non-consumable.
- Over limit alarm and automatic protection.
Accurate Data
- Using 360 air circulation constant temperature technology, the temperature stability is better.
- Equipped with high-precision temperature and humidity sensor, to monitor and record temperature and humidity changes in real time.
- In the test process, the automatic control on flow, temperature and relative humidity is realized, reaching higher accuracy.
- Higher test repeatability of 0.01 cc / (m2day) can be achieved.
High Efficiency Three-cell
- Independent three sets of 50cm2 standard area test cells, in line with parallel sample detection required by standards.
- Three samples can be tested at the same time under the same conditions, with independent tests data.
- In the same test cycle, the number of samples completed increases from 2 to 3.
- Automatic sample clamping, saving time and effort, ensuring consistent clamping force and better sealing.
Intelligent Operation
- The 12 Inch Touch panel of windows system is used for more convenient operation.
- Automatic mode - after inputting test temperature and humidity and clicking one key, the test proceeds automatically.
- New drawer type test cell, one key automatic in-and-out with sound and light reminder.
Safe and Reliable
- Safe Running: Labthinks high-end industrial computer is embedded to eliminate the system failure caused by computer virus and ensure the operation reliability and data storage security.
- Safe Operation: equipped with optical and other intelligent sensors as well as intelligent sound and light alarms to ensure the operation safety.
- Safe Performance: the instrument uses global renowned brand components with stable and reliable performance.
Powerful Functions
- Professional test mode provides flexible and plentiful control functions to meet the needs of scientific research.
- Display oxygen/water vapor transmission rate curve, oxygen/water vapor transmission coefficient curve, temperature curve and humidity curve.
- Wide temperature range, allowing the barrier test at different temperatures (customized).
- The gas purification device independently developed by Labthink can remove trace oxygen in nitrogen, providing oxygen-free carrier gas (optional).
Testing Principle
The pre-treated sample is clamped between the test chambers, oxygen or nitrogen with stable relative humidity flows on one side of the film, and the high purity nitrogen flows on the other side; oxygen or water molecules diffuse through the film into high-purity nitrogen on the other side, and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate can be calculated.
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications
| Applications |
- Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
|
- Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameter/Model |
Model C403H |
| OTR Test Range |
cc/(m2day) (Standard area 50cm2) |
0.01 200 |
| cc/(m2day) (MASK area 5cm2) |
0.2 2000 (Optional) |
| cc/(m2day) (MASK area 1cm2) |
1 10000 (Optional) |
| OTR Resolution |
cc/(m2day) |
0.0001 |
| OTR Repeatability |
cc/(m2day) |
0.01 or 1%, whichever is greater |
| WVTR Test Range |
g/(m2day) (Standard area 50cm2) |
0.005 40 |
| g/(m2day) (MASK area 5cm2) |
0.2 400 (Optional) |
| g/(m2day) (MASK area 1cm2) |
1 2000 (Optional) |
| WVTR Resolution |
g/(m2day) |
0.0001 |
| WVTR Repeatability |
g/(m2day) |
0.005 or 2%, whichever is greater |
| Temperature range |
C |
15 - 50 5 60 (Customized) |
| Temperature fluctuation |
C |
0.05 |
| Humidity range |
%RH (Within standard temperature range) |
0%, 5 - 902% |
| Extended functions |
GP-01 Gas Purification Unit |
Optional |
| DataShieldTM Note3 |
Optional |
| Computer System required by GMP |
Optional |
| CFR21 Part11 |
Optional |
Table 2: Technical Specifications
| Test Chamber |
3 test chambers |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cm) |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard Test Area |
50 cm2 |
| Carrier Gas |
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope) |
| Carrier Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 inch metal tubing |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14either one of twoi14 |
| Instrument Dimensions |
23.6 H x 19.6 W x 27.5 D (60cm 50cm 70cm) |
| Net weight |
220Lbsi14100kgi14 |
Configurations
- Standard Configurations
- Mainframe, tablet, sampler, vacuum grease, polyurethane pipe of 6 mm in diameter
- Optional Parts
- GP-01 Gas Purification Unit, air compressor, CFR21Part11, GMP computer system requirements and DataShieldTM Data ShieldNote3
- Remarks
- 1. The gas supply port of the instrument isI 6 mm polyurethane pipe (pressure 79.7 psi / 550 kPa);
2. Customers will need to prepare for gas supply and distilled water.
Note 1: All the product features are subject to detailed descriptions in the "Technical Parameters.
Note 2: The parameters in the table are measured by professional operators in Labthink Laboratory according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: DataShieldTM Data Shield System provides safe and reliable data application support. The system can be shared by multiple Labthink products. Please purchase separately as needed.
Advanced Testing Principle for Reliable ResultsEmploying both gravimetric and manometric testing methods, this instrument allows for accurate measurement of water vapor, oxygen, and carbon dioxide transmission rates through biodegradable packaging films. Automatic electronic calibration enhances the precision and reliability of each test, while high-sensitivity sensors ensure measurements reflect true material performance.
Intelligent Data Management and User InterfaceThe system features an LCD touchscreen and supports multi-language operation for accessible user experience. Over 500 test results can be stored digitally, with options for data curve analysis and customizable report formats. Automatic operation, specimen conditioning, and microcomputer control streamline the testing process, allowing for consistent workflow in laboratory environments.
Comprehensive Safety and ComplianceEquipped with over-pressure alarms and self-diagnosis functions, the instrument prioritizes operator safety. It meets international testing standards such as ASTM E96, ISO 2528, and GB/T 1037, ensuring your results are aligned with industry best practices, whether for research, quality control, or product development.
FAQ's of Permeability Tests for Packaging Biodegradable Film:
Q: How does the permeability testing instrument evaluate biodegradable film packaging?
A: The instrument utilizes gravimetric and manometric methods to measure the transmission rates of water vapor, oxygen, and carbon dioxide through biodegradable films, providing precise analysis in compliance with ASTM E96, ISO 2528, and GB/T 1037 standards.
Q: What are the benefits of the automatic electronic calibration feature?
A: Automatic electronic calibration minimizes human error and ensures each test is started with optimal accuracy, enhancing the reliability of permeability results throughout repeated measurements.
Q: When is pre-test sample conditioning necessary and how is it supported?
A: Pre-test conditioning is crucial for achieving equilibrium in your samples before measurement, ensuring consistency and accuracy. The instrument supports equilibrium pre-test conditioning within its controlled environment prior to test initiation.
Q: Where can the test results be accessed and in what formats are they available?
A: Test results can be accessed via the LCD touchscreen display and exported digitally through USB or RS-232 interfaces. Reports are available in both digital and printed formats, and data can be stored for over 500 tests.
Q: What is the typical process for conducting a permeability test with this equipment?
A: The process involves conditioning the sample, loading up to three specimens into the chamber, configuring the test parameters via microcomputer control, and starting the test. The system automatically records, analyzes, and outputs the results for further review.
Q: How does the multi-parameter analysis and automation benefit users?
A: Automatic operation with multi-parameter analysis accelerates throughput, reduces manual intervention, and provides comprehensive data, making it easier for users to evaluate packaging film properties efficiently and reliably.
Q: What safety features are incorporated in the instrument?
A: The instrument includes advanced protection such as an over-pressure alarm and self-diagnosis functions, ensuring safe operation during the testing process and prompt detection of any system faults.