Product Description
Product FeaturesNote1
Coulometric Sensor
- Equipped with Labthink latest technology ppb-level coulometric sensor features a wider test range.
- Designed conforming to ASTM D3985, intrinsic standard and no calibration is required.
- Ultra-long service life which three times that of traditional oxygen coulometric sensor.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of air flow, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments.
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click start, the test is fully automated
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with intelligent optics sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
- The gas purifier independently developed by Labthink can remove trace oxygen in nitrogen and provide oxygen-free carrier gas (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, oxygen or air flows on one side of the specimen while a stream of high purity nitrogen flows on the other side. Oxygen molecules permeate through the specimen into the nitrogen side and are carried to the coulometric sensor. The sensor analyzes the oxygen concentration and calculates the oxygen transmission rate and other factors.
Standard
ASTM D3985aASTM F1307aGB/T 19789aGB/T 31354aDIN 53380-3aJIS K7126-2-BaYBB 00082003-2015
Applications
| Applications |
- Films
Oxygen transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others
|
- Sheets
Oxygen transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C206H |
| Test range |
cc/(m2day) (Standard area 50cm2) |
0.02i200 |
| cc/(m2day) (Standard area 50cm2) |
0.02i400000i14Customizei14 |
| cc/(m2day) (MASK area 5cm2) |
0.2i2000i14Optionali14 |
| cc/(m2day) (MASK area 1cm2) |
1i10000i14Optionali14 |
| Resolution |
cc/(m2day) |
0.0001 |
| Repeatability |
cc/(m2day) |
0.02 or 1%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard test temperature rangei14 |
0%i145i902% |
| Additional Functions |
GP-01 Gas Purifier |
Optional |
| DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Gas supply |
99.999% Nitrogena99.5% Oxygeni14 Outside of supply scopei14 |
| Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
GP-01 Gas Purifier, Air compressor, CFR21Part11, GMP Computer System requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI/ 550kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters"
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Precise Multi-Gas Permeability TestingEngineered to measure the permeability of various gases-including oxygen, nitrogen, and carbon dioxide-this system employs the equal pressure method for reliable, repeatable results. Its stainless steel test chamber ensures consistent performance across film types, making it an essential tool for comprehensive material analysis.
Intuitive Operation & Advanced AnalysisFeaturing a touch screen digital display and built-in microcomputer control, the system offers instant access to settings, test progress, and data visualization. The included analysis software provides advanced graphing, customizable reports, and seamless data export via USB, simplifying laboratory workflows.
Robust Safety and AutomationThis instrument delivers automatic calibration and manual adjustment capabilities, paired with over-pressure protection and auto-stop safety functions. Ambient conditions are managed with adjustable temperature and humidity controls, allowing for flexible use in diverse laboratory environments.
FAQ's of Equal Pressure Methods Film Gas Permeability Measuring System:
Q: How does the Equal Pressure Method work in gas permeability testing?
A: The system operates by maintaining equal pressure on both sides of the film specimen, measuring the flow of gases such as oxygen, nitrogen, or carbon dioxide through the material. The microcomputer continually records and analyzes permeability rates, delivering precise, repeatable results.
Q: What types of gases can be tested with this Film Gas Permeability Measuring System?
A: It supports testing with oxygen, nitrogen, carbon dioxide, and other common test gases, catering to a wide variety of film materials and application requirements.
Q: When should automatic and manual calibration be performed?
A: Automatic calibration is recommended for routine tests, ensuring consistent accuracy. Manual calibration can be used for specialized tests, or when fine-tuning is required due to unique specimen characteristics or environmental factors.
Q: Where is this system primarily used?
A: This system is ideally used in film testing laboratories, R&D departments, and quality control environments where precise measurement of gas permeability in films and sheet materials is required.
Q: What is the process for conducting a test using this system?
A: After placing up to three specimens (100 mm 100 mm each) into the stainless steel chamber, users adjust parameters via the touch screen interface. The system runs tests at adjustable speeds, automatically collects and stores data, and provides in-depth analysis with included software.
Q: How does data storage and analysis work on this instrument?
A: All test data are stored internally and can be easily exported via USB. The advanced analysis software allows users to generate graphs and tailored reports, facilitating efficient data interpretation and documentation.
Q: What are the main benefits of using this gas permeability measuring system?
A: Benefits include high precision measurement (accuracy 1% of reading), multi-channel testing for up to three specimens, automated and user-friendly operation, robust safety controls, and advanced data analysis-all contributing to reliable and efficient laboratory workflows.