Product Description
Featuresnote1
Batch Testing
- Equipped with fully-automatic gravity fed sample feeder, which is Labthinks latest scientific and technological achievement that can facilitate continuous testing of 20 to 120 samples.
- A & B dual-channel design can test samples of two different specifications simultaneously.
- The target vacuum degree can be set by the operator to meet testing requirements of different samples.
- Dual methods of vacuum decay and pressure decay provide the flexibility to accommodate various types of package samples.
Precise Data
- Advanced pressure detection technology, using world renowned components for data stability which is not affected by ambient environment.
- Advanced microflow automatic flow control technology that can accurately simulate different sizes of leakage holes without manual adjustment.
- Both the differential pressure transducer and flow meter are traceable to NIST.
- The system can achieve a higher test repeatability of 1um.
Intelligent Control
- 15.6" embedded touch tablet computer with Windows OS.
- New longitudinal interface layout, and easy to use graphic user interface.
- Automatic alarm, automatic capture, automatic collection of unqualified samples.
- The pressure curve is displayed in real time, and the test results are counted automatically.
- Leakage rates are calculated automatically.
- The system is equipped with various sensors as intelligent reminders for safer operation and control.
- Universal printer can be connected for test results output.
- The system features embedded with USB and network ports to facilitate the external access and data transmission of the system, which can be upgraded remotely
Security Compliance
- Verified by compensation and calibration methods.
- The leak tester meets GMP requirements for data traceability and meets the needs of the pharmaceutical industry.
- User operation permission is managed at multiple levels, and the permission content can be configured on demand.
- Electronic signature is designed according to the standard requirements of 21 CFR Part11.
Test Principle
The sample is placed in the sample feeder and automatically delivered into the test cell. The leak rate and other results can be calculated and obtained by analyzing pressure changes measured by the sensor.
Reference Standards
ASTM F2338, YY-T 0681.18, and USP<1207>
Applications
| Basic Applications |
Vials |
Various vial sealing test. |
| Extended Applications |
Ampoule |
Various ampoule bottle sealing tests. |
| Cartridge Bottles |
Various cartridge bottle sealing tests. |
| Injection bottles |
Various injection bottles sealing test. |
Technical Parameters
Table 1: Test Parameters Note2
| ParameterModel |
C690H |
| Testing Range |
um(Reference aperture size USP1207) |
3i8igreat leakage |
| Detection Lower Limit |
um |
3 |
| Resolution |
um |
0.1 |
| Repeatability |
um |
1 |
| Pressure Range |
kPa |
-100i0ii14100 |
| Extended Functions |
21 CFR Part11 |
Optional |
| GMP computer system requirements |
Optional |
Table 2: Technical Specifications
| Testing Cell |
1 set for A group and 1 set for B group |
| Sample Feeder |
1 set for A group and 1 set for B group |
| Sample Size |
I45 mm80mmNote3 |
| Sample Quantity |
20120 pieces |
| Gas Specifications |
Compressed Air (Gas source is provided by the user) |
| Gas Source Pressure |
40.6 PSI / 500 kPa |
| Port Size |
I6 mm Polyurethane tube |
| Dimensions |
33.4 H x 19.6 W x 28.7 D (85cm 50cm 73cm) |
| Power |
120VAC10% 60Hz / 220VAC10% 50Hzi14select one from the twoi14 |
| Net weight |
209Lbs (95kg) |
Product Configuration
| Standard Configuration |
Mainframe, embedded tablet computer, software, flow meter, European vacuum pump, I6 mm polyurethane tube |
| Customization |
Test cell and sample feeder of group A, test cell and sample feeder of group B, negative standard reference sample and positive standard reference sample designed according to sample specifications |
| Optional Parts |
GMP computer system requirements, 21 CFR Part11, air compressor, IQ/OQ/PQ documents |
Note 1: The described product characteristics are subject to the specific annotation of the "Technical Parameters" table.
Note 2: The parameters in the table are measured in the Labthink laboratory by professional operators according to the requirements and conditions of the relevant laboratory environmental standards.
Note 3: Group C test cell can be customized for samples beyond the "Sample Size", but the lower detection limit and test range will change according to the sample size, and the actual delivery shall prevail.
Advanced Leak Detection TechnologyEmploying the pressure decay method, this instrument delivers high sensitivity and reliability for non-destructive leak testing. High-precision differential sensors offer an exceptional accuracy of 0.01 mbar, ensuring even the smallest leaks in electronic modules are detected without causing damage. Each cycle responds in under 2 seconds, providing both speed and confidence in your quality control process.
User-Friendly Operation and Versatile ControlOperate with ease through an intuitive touchscreen interface supporting both English and Chinese languages. The system allows for both automatic and manual modes, giving flexibility for varied testing needs. Automatic and manual calibration options, along with real-time digital LCD display of results, simplify setup and daily operations.
Robust Safety and ComplianceThe instrument is built to the highest standards with over-pressure protection, tool-less maintenance, and a HEPA air filtration system. Its powder-coated steel enclosure is rated at IP54, ensuring durability and safety within challenging laboratory and industrial environments. Complete CE certification guarantees compliance with rigorous international standards.
FAQ's of Non-Destructive Leak Detection Instrument Ensuring Electronic Components Integrity:
Q: How does the pressure decay leak detection method work in this instrument?
A: The instrument introduces dry air at a preset pressure to the specimen and monitors for pressure decay using a high-precision differential pressure sensor. Any leak causes a measurable drop, allowing the system to detect and quantify leaks accurately without damaging the electronic components.
Q: What types of electronic components and sizes can be tested with this device?
A: It can test a wide range of electronic modules, ICs, and microchips with specimen sizes up to 120 mm. The system supports simultaneous testing of one to five specimens, enhancing efficiency in laboratory, production line, and R&D settings.
Q: When should calibration be performed and how can it be done?
A: Calibration can be conducted either automatically or manually, depending on user requirements. Regular calibration-ideally before high-volume testing or after significant maintenance-ensures sustained accuracy. The intuitive UI guides users through the process, making it straightforward and reliable.
Q: Where can the data from test records be stored and exported?
A: The instrument can save up to 5000 test records internally. Data can be exported via RS232 or USB ports, allowing seamless transfer to external computers or laboratory information management systems for further analysis and record-keeping.
Q: What are the safety features included for user protection?
A: Safety is paramount with an over-pressure protection mechanism, audible and visual alarms, and a HEPA air filter that helps maintain a clean testing environment. The IP54 enclosure further shields the device from dust and moisture, enhancing user and equipment safety.
Q: How is the instrument maintained and serviced?
A: Routine maintenance is simplified by the tool-less access design, enabling quick servicing and filter changes. Minimal tools are required, reducing equipment downtime and ensuring continued availability for testing operations.
Q: What benefits does this instrument offer compared to other leak detection methods?
A: This non-destructive instrument provides precise, rapid results with high repeatability, eliminates component damage risks, supports advanced data management, and features comprehensive safety and compliance credentials. It enhances reliability in quality assurance while supporting both production and R&D requirements.