Product Description
Product FeaturesNote1
Coulometric Sensor
- Equipped with Labthink latest technology ppb-level coulometric sensor features a wider test range.
- Designed conforming to ASTM D3985, intrinsic standard and no calibration is required.
- Ultra-long service life which three times that of traditional oxygen coulometric sensor.
- Embedded with over-range warning and automatic protection function.
Accurate data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- The test chamber is equipped with a high-precision humidity sensor to monitor and record humidity changes in real time.
- The control of air flow, temperature and relative humidity is automated to realize higher accuracy.
High Efficiency
- Six independent test cells with a standard area of 50cm2, three times the number of test cells in traditional oxygen permeability testing instruments.
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Within the same time duration, the number of tested specimens is increased from 2 to 6.
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click start, the test is fully automated
- Intelligent test chamber hood automatically opens and closes with sound and light alert.
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with intelligent optics sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides oxygen transmission rate curve, oxygen transmission coefficient curve, temperature curve, and humidity curve.
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Ultra-wide temperature range to meet the barrier test under different extreme temperatures (optional).
- The gas purifier independently developed by Labthink can remove trace oxygen in nitrogen and provide oxygen-free carrier gas (optional).
Test Principle
The pre-conditioned specimen is clamped in the test cell, oxygen or air flows on one side of the specimen while a stream of high purity nitrogen flows on the other side. Oxygen molecules permeate through the specimen into the nitrogen side and are carried to the coulometric sensor. The sensor analyzes the oxygen concentration and calculates the oxygen transmission rate and other factors.
Standard
ASTM D3985aASTM F1307aGB/T 19789aGB/T 31354aDIN 53380-3aJIS K7126-2-BaYBB 00082003-2015
Applications
| Applications |
- Films
Oxygen transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others
|
- Sheets
Oxygen transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C206H |
| Test range |
cc/(m2day) (Standard area 50cm2) |
0.02i200 |
| cc/(m2day) (Standard area 50cm2) |
0.02i400000i14Customizei14 |
| cc/(m2day) (MASK area 5cm2) |
0.2i2000i14Optionali14 |
| cc/(m2day) (MASK area 1cm2) |
1i10000i14Optionali14 |
| Resolution |
cc/(m2day) |
0.0001 |
| Repeatability |
cc/(m2day) |
0.02 or 1%i14take the greater |
| Test Temperature |
|
15i50i145i60i14Optionali14 |
| Temperature fluctuation |
|
0.15 |
| Test Humidity |
%RHi14Within standard test temperature rangei14 |
0%i145i902% |
| Additional Functions |
GP-01 Gas Purifier |
Optional |
| DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.4 x 4.4i1411.2cm11.2cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
50cm2 |
| Gas supply |
99.999% Nitrogena99.5% Oxygeni14 Outside of supply scopei14 |
| Gas Pressure |
40.6 PSI / 280 kPa |
| Port Size |
1/8 Metal tube |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz / 220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
GP-01 Gas Purifier, Air compressor, CFR21Part11, GMP Computer System requirement, DataShieldTM note3 |
| Note |
The gas supply port of the instrument is I6 mm PU tubingi14pressure79.7 PSI/ 550kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters"
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Accurate & Reliable Oxygen Barrier AssessmentThis equipment delivers highly accurate measurements (0.01 cc/m2day) with a resolution as fine as 0.001 cc/m2day. Offering a wide measurement range (0.01-100 cc/m2day), it ensures both low and high permeability materials are tested with confidence. High-throughput capability, universal clamp type holders, and user-friendly automation support efficient, repeatable assessments.
User-Friendly Digital Operation & Data ManagementWith its color touch screen display, automatic calibration, and intuitive Windows-based software, the tester streamlines daily testing workflows. Test results can be downloaded via USB, stored locally, or uploaded for cloud-based access. Seamless data management and alarm systems enhance laboratory productivity and traceability.
Quality Assurance & Compliance for Food PackagingFully compliant with ASTM D3985 and ISO 15105-2, the tester underpins food packaging analysis by ensuring sample conditioning and controlled testing environments. The robust stainless steel chamber and integrated environmental controls guarantee reliable, reproducible results-a key benefit for manufacturers, suppliers, and exporters seeking to meet stringent quality standards.
FAQ's of Evaluating Oxygen Barrier Properties of Plastic Cups Ensuring Food Shelf life:
Q: How does the Oxygen Permeability Tester evaluate the oxygen barrier properties of plastic cups?
A: The tester employs the coulometric sensor method to measure the oxygen transmission rate (OTR) of plastic cup samples. A controlled stream of oxygen passes through the sample, and the amount permeating is precisely detected by a coulometric sensor, ensuring reliable assessment of material barrier properties.
Q: What standards does this equipment meet for food packaging analysis?
A: This Oxygen Permeability Tester complies with ASTM D3985 and ISO 15105-2 standards. These certifications guarantee that the measurements and methods are internationally recognized for food packaging and quality control applications.
Q: When should I use sample conditioning for my tests?
A: Sample conditioning is recommended before every test to ensure your plastic cup specimens reach specified humidity and temperature conditions (40-90% RH, 15-50C). This process helps provide accurate and repeatable measurements that reflect real-world storage and usage environments.
Q: Where can test results be stored and accessed?
A: Test results can be accessed directly on the device's color touch screen, downloaded via USB, or stored locally. Additionally, the system is cloud-enabled, allowing users to save and retrieve data remotely for seamless laboratory and quality documentation.
Q: What is the process of calibration, and how often is it required?
A: Calibration is handled automatically by the integrated system, utilizing built-in standard gas calibration. This ensures ongoing measurement accuracy and eliminates the need for frequent manual calibration, reducing downtime and potential errors.
Q: How is the tester operated, and what are its software compatibility features?
A: Controlled through a color touch screen and supported by Windows-based software, the tester offers an intuitive interface and remote data download capabilities through USB. The microprocessor control guarantees consistent, automated operation throughout the measurement cycle.
Q: What are the main benefits of using this Oxygen Permeability Tester?
A: The tester offers high measurement accuracy, rapid response time (<30 s), integrated sample conditioning, efficient storage options (cloud and local), automatic calibration, and full compliance with industry standards. It streamlines laboratory workflows and supports robust quality assurance for food packaging manufacturers and suppliers.