Product Description
Featuresnote1
Batch Testing
- Equipped with fully-automatic gravity fed sample feeder, which is Labthinks latest scientific and technological achievement that can facilitate continuous testing of 20 to 120 samples.
- A & B dual-channel design can test samples of two different specifications simultaneously.
- The target vacuum degree can be set by the operator to meet testing requirements of different samples.
- Dual methods of vacuum decay and pressure decay provide the flexibility to accommodate various types of package samples.
Precise Data
- Advanced pressure detection technology, using world renowned components for data stability which is not affected by ambient environment.
- Advanced microflow automatic flow control technology that can accurately simulate different sizes of leakage holes without manual adjustment.
- Both the differential pressure transducer and flow meter are traceable to NIST.
- The system can achieve a higher test repeatability of 1um.
Intelligent Control
- 15.6" embedded touch tablet computer with Windows OS.
- New longitudinal interface layout, and easy to use graphic user interface.
- Automatic alarm, automatic capture, automatic collection of unqualified samples.
- The pressure curve is displayed in real time, and the test results are counted automatically.
- Leakage rates are calculated automatically.
- The system is equipped with various sensors as intelligent reminders for safer operation and control.
- Universal printer can be connected for test results output.
- The system features embedded with USB and network ports to facilitate the external access and data transmission of the system, which can be upgraded remotely
Security Compliance
- Verified by compensation and calibration methods.
- The leak tester meets GMP requirements for data traceability and meets the needs of the pharmaceutical industry.
- User operation permission is managed at multiple levels, and the permission content can be configured on demand.
- Electronic signature is designed according to the standard requirements of 21 CFR Part11.
Test Principle
The sample is placed in the sample feeder and automatically delivered into the test cell. The leak rate and other results can be calculated and obtained by analyzing pressure changes measured by the sensor.
Reference Standards
ASTM F2338, YY-T 0681.18, and USP<1207>
Applications
| Basic Applications |
Vials |
Various vial sealing test. |
| Extended Applications |
Ampoule |
Various ampoule bottle sealing tests. |
| Cartridge Bottles |
Various cartridge bottle sealing tests. |
| Injection bottles |
Various injection bottles sealing test. |
Technical Parameters
Table 1: Test Parameters Note2
| ParameterModel |
C690H |
| Testing Range |
um(Reference aperture size USP1207) |
3i8igreat leakage |
| Detection Lower Limit |
um |
3 |
| Resolution |
um |
0.1 |
| Repeatability |
um |
1 |
| Pressure Range |
kPa |
-100i0ii14100 |
| Extended Functions |
21 CFR Part11 |
Optional |
| GMP computer system requirements |
Optional |
Table 2: Technical Specifications
| Testing Cell |
1 set for A group and 1 set for B group |
| Sample Feeder |
1 set for A group and 1 set for B group |
| Sample Size |
I45 mm80mmNote3 |
| Sample Quantity |
20120 pieces |
| Gas Specifications |
Compressed Air (Gas source is provided by the user) |
| Gas Source Pressure |
40.6 PSI / 500 kPa |
| Port Size |
I6 mm Polyurethane tube |
| Dimensions |
33.4 H x 19.6 W x 28.7 D (85cm 50cm 73cm) |
| Power |
120VAC10% 60Hz / 220VAC10% 50Hzi14select one from the twoi14 |
| Net weight |
209Lbs (95kg) |
Product Configuration
| Standard Configuration |
Mainframe, embedded tablet computer, software, flow meter, European vacuum pump, I6 mm polyurethane tube |
| Customization |
Test cell and sample feeder of group A, test cell and sample feeder of group B, negative standard reference sample and positive standard reference sample designed according to sample specifications |
| Optional Parts |
GMP computer system requirements, 21 CFR Part11, air compressor, IQ/OQ/PQ documents |
Note 1: The described product characteristics are subject to the specific annotation of the "Technical Parameters" table.
Note 2: The parameters in the table are measured in the Labthink laboratory by professional operators according to the requirements and conditions of the relevant laboratory environmental standards.
Note 3: Group C test cell can be customized for samples beyond the "Sample Size", but the lower detection limit and test range will change according to the sample size, and the actual delivery shall prevail.
Accurate, Non-Destructive Seal TestingAchieve precision quality control with rapid, non-destructive testing methods that preserve packaging. This machine detects minute leaks as small as 0.05 mm in diameter, ensuring integrity across a spectrum of packaging materials in pharmaceutical, food, and industrial sectors. Its robust automation and sensitivity redefine seal verification and compliance.
User-Friendly Digital InterfaceThe tester features a large touchscreen digital display, intuitive microprocessor-based control, and supports both English and multiple languages. Customizable test protocols, automatic test sequences, and rapid response times streamline operation, making routine inspection efficient and user-friendly for all levels of operators.
Compliance and TraceabilityFully aligned with ISO 11607 and ASTM F2338 standards, this instrument guarantees compliant testing processes. Internal memory for 1,000 records, plus USB/RS232 interfaces, secure all data for audits, traceability, and regulatory requirements. The system's alarm functions provide immediate notification of seal failures to ensure prompt corrective action.
FAQ's of Packaging Seal Integrity Testing with Non-Destructive Methods:
Q: How does the non-destructive package seal integrity tester detect leaks?
A: This equipment employs vacuum decay and pressure decay test modes to identify leaks as small as 0.05 mm in diameter. By measuring changes in pressure without damaging the packaging, the tester quickly highlights areas with compromised seal integrity.
Q: What are the main benefits of using this tester for routine packaging inspection?
A: The tester offers rapid, non-destructive assessment, ensuring packaging remains intact post-inspection. Its high sensitivity, compliance with ISO and ASTM standards, automation, and digital data management contribute to reliable, efficient quality control and improved traceability.
Q: When should this seal integrity tester be used in the production process?
A: This machine is ideal for use during routine quality assurance inspections, post-packaging, and prior to product release. It ensures that any leaks are detected early, reducing the risk of product recalls and regulatory non-compliance.
Q: Where is this equipment best installed for optimal operation?
A: With its cleanroom compatibility and compact benchtop design, the tester is best suited for QC laboratories, packing lines, or cleanroom environments in pharmaceutical, food, and medical device manufacturing facilities.
Q: What features ensure ease of operation and user convenience?
A: The system provides a touchscreen digital display, microprocessor-based automatic control, customizable test protocols, audible and visual alarms, as well as English and multi-language support. Quick response and digital record management enhance everyday usability.
Q: How does the data storage and alarm system support process monitoring?
A: Internal memory securely stores up to 1,000 test records, which can be easily accessed or exported via USB/RS232 interfaces. The alarm system delivers audible and visual alerts for immediate reporting of test failures, enabling fast corrective action.
Q: What support options are available for maintenance and troubleshooting?
A: Comprehensive service support includes both on-site and remote diagnostics, ensuring minimal downtime and continued precision performance throughout the lifespan of the equipment.