Product Description
Product FeaturesNote1
Accurate Data
- Brand new dome design test chamber and 360 air circulation constant temperature technology ensures better temperature stability.
- Imported high-precision vacuum sensor can accurately test ultra-high barrier materials with 0.01i0.09cm3/ m324h0.1MPa test range.
- Imported pneumatic control system has a low failure rate and an ultra-long service life, ensuring good overall system air tightness and high test accuracy.
- Imported vacuum pump can reach 0.2Pa extreme pressure with increased vacuum rate.
- The vacuum pump is controlled automatically by the system without manual opening and closing, increases efficiency and saves labor.
- Labthinks unique high-precision automatic pressure compensation technology in the test process realizes a constant pressure difference in the high-pressure chamber, and the pressure fluctuation is less than 0.2 kPa.
- The pressure in high-pressure chamber can be set within the range of 10kPa210kPa, the pressure can be maintained accurately and stably by the system.
High Efficiency
- Six independent test cells with standard area, three times the number of test cells in traditional oxygen permeability testing instruments.
- Six specimens can be tested simultaneously under the same testing condition, delivering independent test result.
- Medium and low barrier materials, test time <4 hours (including vacuumize time).
- High barrier materials, test time <4 hours (including vacuumize time).
- Automatic specimen clamping saves time and effort. The clamping force is consistent, resulting in better air tightness.
Intelligent control
- 12 touch-screen tablet powered by WindowsTM 10 operating system makes the operation simpler and more convenient.
- Automatic test mode requires only inputting temperature and humidity, one click start, the test is fully automated.
- Intelligent test chamber hood automatically opens and closes with sound and light alert
Safe and reliable
- System security -- Built-in Labthink's unique high-end industrial computer prevents system failures caused by computer viruses, ensures operational reliability and data storage security.
- Operation safety-- Equipped with intelligent optics sensors which give sound and light alert to ensure safe operation.
- Performance reliabilitythe instrument adopts components of global renowned brands, to ensure stable and reliable performance.
Space saving
- The width of the instrument is only 1/3 of the traditional six-cell instrument, saving space for a laboratory.
Powerful functions
- Professional test mode provides flexible and diverse control options to meet various needs of scientific research.
- The system provides gas transmission rate curve, gas transmission coefficient curve, and temperature curve.
- Ultra-wide test temperature range to enable the barrier tests under various extreme temperature (Optional).
- Ultra-wide test range to meet the barrier test of various materials (customize).
- Tests of hazardous gases such as H2aCH4 are applicable (Customize).
Test Principle
The pre-conditioned specimen is mounted in the test cell as to form a sealed barrier between two chambers. The lower-pressure chamber (lower chamber) is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber (upper chamber) and a constant pressure (adjustable) difference is generated between the two chambers. The gas permeates through the specimen from higher pressure side into the lower pressure side. The gas permeability and other barrier properties of the specimen can be obtained by analyzing the pressure changes in the lower chamber.
Standard
ISO 15105-1aISO 2556aGB/T 1038-2000aASTM D1434aJIS K7126-1aYBB 00082003
Applications
| Applications |
- Films
Gas transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others
|
- Sheets
Gas transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials
|
Technical Specifications
Table 1: Test Parameters Note2
| Parameters/Model |
C106H |
| Test range |
cm3/m224h0.1MPa |
0.01i50,000 |
| Resolution |
cm3/m224h0.1MPa |
0.001 |
| Test temperature |
|
15i50i145i60i14Optionali14 |
| Temperature resolution |
|
0.01 |
| Temperature fluctuation |
|
0.15 |
| Test humidity |
%RHi14Within standard test temperature rangei14 |
0%i145i902%i14Customizei14 |
| Vacuum Resolution |
Pa |
0.01 |
| Vacuum Accuracy |
Pa |
0.2%i141%-100% of Full Scalei14 |
| Vacuum Degree of Test Chamber |
Pa |
10 |
| Additional Functions |
DataShieldTM note3 |
Optional |
| GMP Computer System requirement |
Optional |
| CFR21Part11 |
Optional |
Table 2: Technical Specifications
| Test Cell |
6 Cells |
| Specimen Size |
4.6 x 4.6i1411.7cm11.7cmi14 |
| Specimen Thickness |
120 Mili143mmi14 |
| Standard test area |
38.48cm2 |
| Test Gas |
O2aN2aCO2 etc.i14Outside of supply scopei14 |
| Test Pressure |
10 kPa210 kPai14Can be set betweeni14 |
| Pressure Fluctuation |
0.2 kPa |
| Gas Pressure |
79.7 PSI / 550 kPa |
| Port Size |
I6 mm PU tubing |
| Instrument Dimension |
23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
| Power Supply |
120VAC10% 60Hz/220VAC10% 50Hzi14one of twoi14 |
| Net Weight |
220Lbsi14100kgi14 |
Table 3: Product Configuration
| Standard Configuration |
Instrument mainframe, tablet, vacuum pump (UK), sampler, vacuum grease, I6 mm PU tubing |
| Optional Parts |
Air compressor, CFR21Part11, GMP Computer System requirement, DataShieldTM note3 |
| Note |
The gas compressor port of the instrument is I6 mm PU tubingi14pressure79.7 PSI/ 550kPai14, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters" "Table 1: Test Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
Reliable Barrier Property AnalysisThis instrument utilizes a high-sensitivity pressure sensor to evaluate the barrier effectiveness of coated papers and packaging films. Its digital LCD provides fast, precise readings, ensuring dependable results for various research and quality control applications. The programmer-friendly microprocessor streamlines operations, allowing automated, repeatable test sequences.
Advanced Environmental ControlIntegrated temperature and humidity control maintains optimal testing conditions, accommodating a humidity range of 40-90% RH and temperatures from room temperature to 50C. Environmental consistency ensures that each test run is comparable and valid, vital for accurate assessments of barrier properties across multiple samples.
Efficient, Safe, and Intuitive OperationThe tester supports simultaneous testing of three specimens, empowering laboratories to improve workflow efficiency. Safety is paramount, with overload protection and an emergency stop switch. Data management is simplified with large storage capacity, intuitive user interface, and seamless PC connectivity for deeper data analysis and reporting.
FAQ's of Coated Paper Barrier Properties Tester:
Q: How does the Coated Paper Barrier Properties Tester measure barrier properties?
A: The tester uses a high-sensitivity pressure sensor to detect and record changes in the barrier property parameters, such as permeability and resistance, across coated paper or packaging films. Measurements are quickly displayed on its digital LCD for easy reading.
Q: What is the process for testing multiple paper samples at once?
A: The instrument allows simultaneous testing of up to three specimens, each measuring 100 mm 100 mm. This enhances throughput, as all three are tested under identical environmental and operational conditions using programmable test cycles.
Q: When should I calibrate the device, and is calibration difficult?
A: Calibration can be performed automatically or manually, depending on your preference. It is recommended to calibrate the tester before each new batch of measurements or after relocation, ensuring reliable and accurate results every time.
Q: Where is this barrier properties tester typically used?
A: This equipment is primarily used in laboratory settings, including quality assurance and research labs in paper manufacturing, packaging, and related industries. Its bench-top design supports convenient integration into standard lab workflows.
Q: What safety features are included during operation?
A: For operator and equipment safety, the tester includes overload protection and an emergency stop switch. These features protect against unexpected issues during testing and allow quick halting in emergency cases.
Q: How is data managed and analyzed after testing?
A: The tester can store up to 500 test results internally. Data can be transferred via RS232 or USB to a computer, where bundled PC software enables comprehensive analysis and automatic generation of detailed test reports.
Q: What are the main benefits of using this specific barrier properties tester model?
A: Users benefit from high precision and speed, the ability to test multiple samples simultaneously, robust environmental control, automated reporting capabilities, and enhanced operator safety-all supporting fast, reliable, and repeatable material testing.